Measurement of sirolimus concentrations in human blood using an automated electrochemiluminescence immunoassay (ECLIA): a multicenter evaluation.
Clicks: 409
ID: 72774
2018
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
Steady Performance
71.3
/100
409 views
277 readers
Trending
AI Quality Assessment
Not analyzed
Readership in this journal
SteadyRanked #4 of 31 articles by views in Clinical chemistry and laboratory medicine
Most read
Least read
Bar heights use a square-root scale.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
Therapeutic drug monitoring (TDM) of sirolimus is essential in transplant recipients. We evaluated the performance of a new electrochemiluminescence immunoassay (ECLIA) for measuring sirolimus concentrations in whole blood at five European laboratories.Study assessments included repeatability, intermediate precision and functional sensitivity (concentration at coefficient of variation [CV] of 20%) experiments. Method comparisons with liquid chromatography-tandem mass spectrometry (LC-MS/MS; reference method) and two immunoassays (chemiluminescent microparticle immunoassay [CMIA] and antibody-conjugated magnetic immunoassay [ACMIA]) were performed using native samples from patients with kidney transplants.Imprecision testing CVs were ≤6.4% and ≤10.7% across the sirolimus concentration range for both repeatability and intermediate precision, respectively. The ECLIA showed excellent functional sensitivity: the CV did not reach 20%; the CV at the assay's limit of quantitation (1.5 μg/L) was 7.0%. Agreement between the ECLIA and LC-MS/MS using native kidney samples was close, with weighted Deming regression analysis yielding a slope of 1.05, an intercept of 0.154 μg/L and a Pearson's correlation coefficient (r) of 0.94, while Bland-Altman analysis showed a combined mean bias of 0.41 μg/L (±2 standard deviation [SD], -1.96 to 2.68). The ECLIA also showed good correlation with the two other immunoassays: the CMIA (slope=0.91, intercept=0.112 μg/L and r=0.89) and the ACMIA (slope=0.99, intercept=0.319 μg/L and r=0.97).The ECLIA showed good precision, functional sensitivity and agreement with other methods of sirolimus measurement used in clinical practice, suggesting that the assay is suitable for TDM in transplant recipients and provides an alternative to LC-MS/MS.
| Reference Key |
stove2018measurementclinical
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | Stove, Veronique;Ramos, Pedro Alía;Wallemacq, Pierre;Vogeser, Michael;Schuetzenmeister, Andre;Schmiedel, Christian;Shipkova, Maria; |
| Journal | Clinical chemistry and laboratory medicine |
| Year | 2018 |
| DOI |
10.1515/cclm-2017-0583
|
| URL | |
| Keywords |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.