6 MV small field output factors: intra-/inter-machine comparison and implementation of TRS-483 using various detectors and several linear accelerators.

Clicks: 257
ID: 47432
2019
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Steady

Ranked #30 of 56 articles by views in Medical physics

Most read Least read

Bar heights use a square-root scale.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
To investigate the applicability of output correction-factors reported in TRS-483 on 6 MV small-field detector-reading ratios using four solid-state detectors. Also, to investigate variations in 6 MV small-field output factors (OF) among nominally-matched linear accelerators (linacs).The TRS-483 Code of Practice (CoP) introduced and provided output correction-factors to be applied to measured detector-reading ratios to obtain OFs for several small-field detectors. Detector-readings for 0.5 cm x 0.5 cm to 8 cm x 8 cm fields were measured and normalized to that of 10 cm x 10 cm field giving the detector-reading ratios. Three silicon diodes: IBA PFD, IBA EFD (IBA, Schwarzenbruck, Germany), PTW T60017 and one microdiamond: PTW T60019 (PTW, Freiburg, Germany) were used. Output correction-factors from the CoP were applied to measured detector-reading ratios. Measurements were performed on six Clinac and six TrueBeam linacs (Varian Medical Systems, Palo Alto, USA). An investigation of the relationship between the size of small fields and corresponding detector-reading ratio among the linacs was performed by measuring lateral dose-profiles for 0.5 cm x 0.5 cm fields to determine the Full Width Half Maximum (FWHM). The relationship between the linacs' focal-spot size and the small-field detector-reading ratio was investigated by measuring 10 cm x 10 cm lateral dose-profiles and determining the penumbra width reflecting the focal-spot size. Measurement geometry were as follows: gantry angle = 0°, collimator angle = 0°, source-to surface distance (SSD) = 90 cm and depth in water = 10 cm.For a given linac and 0.5 cm x 0.5 cm field, the deviations in detector-reading ratios among the detectors were 9-15 % for the Clinacs and 4-5 % for the TrueBeams. Use of output correction-factors reduced these deviations to 6-12 % and 3-4 % respectively. For field-sizes equal to or larger than 0.8 cm x 0.8 cm, the deviations were corrected to 1 % using output correction-factors for both Clinacs and TrueBeams. For a given detector and 0.5 cm x 0.5 cm field, the deviations in detector-reading ratios among the linacs were 11-17 % for the Clinacs and 5-6 % for the TrueBeams. For 1 cm x 1 cm the deviations were 1-2 % for Clinacs and 1 % for TrueBeams. For field-sizes larger than 1 cm x 1 cm the deviations were within 1 % for both Clinacs and TrueBeams. No relationship between FWHMs and detector-reading ratios for 0.5 cm x 0.5 cm was observed. For Clinacs, larger 10 cm x 10 cm penumbra width yielded lower 0.5 cm x 0.5 cm detector-reading ratio indicating an effect of the focal spot size. For TrueBeams, the spread of penumbra widths was lower compared to Clinacs and no similar relationship was observed.Output correction-factors from the TRS-483 CoP are not sufficient for accurate determination of OF for 0.5 cm x 0.5 cm fields but are applicable for 0.8 cm x 0.8 cm to 8 cm x 8 cm fields. Nominally-matched Clinacs and TrueBeams show large differences in detector-reading ratios for fields smaller than 1 cm x 1 cm.
Reference Key
ghazal20196medical Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors Ghazal, Mohammed;Westermark, Mathias;Kaveckyte, Vaiva;Carlsson-Tedgren, Åsa;Benmakhlouf, Hamza;
Journal Medical physics
Year 2019
DOI
10.1002/mp.13830
URL
Keywords

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.