Injection Molded Liquid Crystal Polymer Package for Chronic Active Implantable Devices with Application to an Optogenetic Stimulator.
Clicks: 245
ID: 38050
2019
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
Steady Performance
30.0
/100
245 views
31 readers
AI Quality Assessment
Not analyzed
Readership in this journal
SteadyRanked #13 of 16 articles by views in ieee transactions on bio-medical engineering
Most read
Least read
Bar heights use a square-root scale.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
Implanted electronics require protection from the body's fluids to avoid moisture induced failure. This study presents an injection molded liquid crystal polymer (LCP) package to protect active implantable devices for chronic applications, such as in optogenetic research. The technology is applied and assessed through a custom package for a fully implantable optogenetic stimulation system, built on a versatile telemetry system that can incorporate additional stimulating and recording channels. An adapted quasi-steady state model predicts the lifetime of an enclosure, where the definition of the lifetime is the time before the internal relative humidity (RH) reaches a time constant, or 63 %RH, a conservative limit to minimize the risk of corrosion. The lifetime of the LCP optogenetic device is 94 days, and can be extended to 326 days with the inclusion of 5 % w/v silica gel desiccant. Samples of the LCP optogenetic device containing humidity sensors testing in saline at 38 °C support the RH change predictions. Desiccants inside the implant enclosure can store permeating moisture and prolong the life expectancy of LCP-based implants to years or decades. The results of this study demonstrates the feasibility of providing reliable protection for chronic optogenetic implants, and the technology can be transferred to other applications as an easily-manufactured, cost-effective, radiofrequency compatible alternative to hermetic packaging for chronic studies.
| Reference Key |
au2019injectionieee
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | Au, Shun Long Cyril;Chen, Fu-Yu Beverly;Budgett, David;Malpas, Simon C;Guild, Sarah-Jane;McCormick, Daniel; |
| Journal | ieee transactions on bio-medical engineering |
| Year | 2019 |
| DOI |
10.1109/TBME.2019.2936577
|
| URL | |
| Keywords | Keywords not found |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.