Nocturnal Blood Pressure Dipping in Adults: A Comparison Between Cuffless and Ambulatory Blood Pressure Monitoring
Clicks: 9
ID: 329476
2026
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
Emerging Content
2.4
/100
9 views
8 readers
AI Quality Assessment
Not analyzed
Readership in this journal
EmergingRanked #68 of 78 articles by views in American Journal of Hypertension
Most read
Least read
Bar heights use a square-root scale.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
BACKGROUND: Cuffless blood pressure (BP) devices have demonstrated potential in estimating daytime BP but have not been able to predict nocturnal BP precisely, creating difficulties when the intended use is ambulatory BP monitoring (ABPM). We evaluated the ability of a photoplethysmography-based cuffless BP device, with the intended use 24-hour ABPM, to estimate nocturnal BP compared to a cuff-based oscillometric device. METHODS: Participants underwent 24-hour ABPM with the cuffless device placed on the non-dominant upper arm and a validated cuff-based oscillometric device on the dominant arm. The difference between mean awake and asleep BP (awake-asleep BP change) was calculated for both devices and used in the analyses. The statistical analyses were guided by the Awake/Asleep test (European Society of Hypertension 2023 Recommendations), however, there were methodological deviances from these recommendations. Most importantly, we used only cuffless BP values that were obtained simultaneously with the reference device. RESULTS: Data from 50 participants, aged 18-80 years, with mean (standard deviation; SD) inclusion office systolic/diastolic BP (SBP/DBP) 120.1 (11.5)/75.3 (7.9), were analyzed. The cuffless device achieved a mean difference [95% limits of agreement] of 1.6 [-8.0, 11.2] mmHg for SBP and 1.2 [-6.7, 9.1] mmHg for DBP in awake-asleep BP changes compared to the reference. Pearson correlation coefficients (95% confidence interval) for SBP/DBP were 0.75 (0.59, 0.85)/0.74 (0.58, 0.84). Our population had an overrepresentation of normal and low BP. CONCLUSION: Although the cuffless device shows promise in successfully tracking nocturnal BP, full validation is required before clinical use.
| Reference Key |
openalex_W7214001606
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | C Hove, T M Seeberg, K G Bøtker-Rasmussen, E Westgaard, S Heimark, B Waldum-Grevbo, A C K Larstorp |
| Journal | American Journal of Hypertension |
| Year | 2026 |
| DOI |
10.1093/ajh/hpag115
|