Comparison of Histological Effects of LigaSure and Bipolar Cautery on the Submandibular Gland in a Rat Model
Clicks: 1
ID: 318765
2026
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
0.0
/100
1 views
0 readers
AI Quality Assessment
Not analyzed
Readership in this journal
Ranked #72 of 73 articles by views in Aesthetic surgery journal
Most read
Least read
Bar heights use a square-root scale.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
BACKGROUND: Submandibular gland prominence is increasingly recognized as a limiting factor in achieving optimal cervicomental contour. Although energy-based devices are widely used during glandular procedures, their tissue-level effects remain insufficiently defined. OBJECTIVES: To compare the histological impact of a vessel-sealing system and conventional bipolar electrocautery on submandibular gland tissue in an experimental model. METHODS: Twenty-four male Wistar rats were randomly assigned to three groups (LigaSure, bipolar electrocautery, and cold dissection control; n=8 per group). Partial glandular excision was performed under standardized conditions. Tissue specimens were processed for histological analysis using hematoxylin and eosin staining. Edema, hemorrhage, vascular injury, necrosis, inflammation, and ductal narrowing were assessed using a semi-quantitative scoring system. Non-parametric statistical methods were applied for intergroup comparisons. RESULTS: Both energy-based modalities produced significantly greater tissue alterations compared with controls (p<0.05). Bipolar electrocautery demonstrated higher levels of inflammation, edema, hemorrhage, and necrosis relative to LigaSure (p<0.05), while vascular injury showed no statistically significant difference. In contrast, ductal narrowing was significantly more pronounced in the LigaSure group (p<0.05). Overall, the most marked differences between modalities were observed in the inflammatory response and ductal changes. CONCLUSIONS: Distinct patterns of tissue injury are associated with different energy modalities. LigaSure appears to limit collateral tissue damage while exerting a greater effect on ductal structures, whereas bipolar electrocautery is associated with more extensive inflammatory and hemorrhagic changes. These findings provide experimental insight that may inform energy device selection in submandibular gland surgery, although further studies are needed to clarify their clinical implications.
| Reference Key |
openalex_W7165856867
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | Serhat Şibar, Muzaffer Duran, Sami Can Yeşilırmak, Elham Bahador Zırh, Selim Zırh, Mine Farımaz |
| Journal | Aesthetic surgery journal |
| Year | 2026 |
| DOI |
10.1093/asj/sjag127
|
| URL | |
| Keywords | Keywords not found |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.