Evaluation of the Image Quality of Ultra-Low-Dose Paranasal Sinus Computed Tomography Scans
Clicks: 200
ID: 276958
2023
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
Steady Performance
30.0
/100
200 views
37 readers
AI Quality Assessment
Not analyzed
Readership in this journal
SteadyRanked #6 of 18 articles by views in European Journal of Therapeutics
Most read
Least read
Bar heights use a square-root scale.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
Objective: We aimed to investigate the image quality of paranasal sinus computed tomography (CT) scans obtained with the “Advanced intelligent Clear-IQ Engine” (AICE) software and ultra-low dose parameters in patients with prediagnosed rhinitis, sinusitis or nasal septum deviation.
Methods: The first 50 patients (31 men and 19 women, aged 18-70 years) who agreed to participate in our prospectively planned study were included in the study. Imaging of the patients was performed with a 160-slice multidetector CT device Canon Aquilion Prime SP (Canon Medical Systems, Otawara, Japan). Tube voltage of 80 kV and tube current of 40 mA were chosen. The effective radiation dose (ED) was calculated based on the “dose-length product” (DLP) information of each patient. Evaluation of images was made on reformat images made with “AiCE Bone” filter in the axial and coronal planes. All images were double-blindly evaluated by 2 radiologists out of 5 points (5 = very good; 1 = not diagnostic). The visibility of 17 bilateral important anatomical landmarks examined on paranasal sinus CT was scored (0 = not visible; 2 = completely visible). In the evaluation of images, the agreement between radiologists was evaluated with Kappa statistics.
Results: The median ED value of the CT protocol was 0.015±0.001 mSv [range 0.012-0.017 mSv]. The overall image quality of the images in our study was interpreted as 4 (good) or 3 (moderate) for all patients. Inter-research agreement was 92%. Most of the anatomical landmarks evaluated were completely visible. Six (bilateral of three) anatomical landmarks could be partially visible in some patients.
Conclusion: Combining the ultra-low dose CT protocol and the AiCE reconstruction algorithm makes it possible to achieve very low ED values in paranasal sinus CT scans and still obtain sufficient image quality in indications such as septum deviation and sinusitis.
| Reference Key |
Akşamoğlu2023europeanEvaluation
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | Melih Akşamoğlu;Mehmet Sait Menzilcioğlu; |
| Journal | European Journal of Therapeutics |
| Year | 2023 |
| DOI |
1131
|
| URL | |
| Keywords |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.