An Evaluation of Agreement of Breathing Rates Measured by a Novel Device, Manual Counting, and Other Techniques Used in Clinical Practice: Protocol for the Observational VENTILATE Study
Clicks: 132
ID: 273682
2020
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
Emerging Content
30.0
/100
132 views
24 readers
AI Quality Assessment
Not analyzed
Readership in this journal
EmergingRanked #94 of 113 articles by views in jmir research protocols
Most read
Least read
Bar heights use a square-root scale.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
Background: Respiratory rate (RR) is the most sensitive physiological observation to predict clinical deterioration on hospital wards, and poor clinical monitoring has been highlighted as a primary contributor to avoidable mortality. Patients in intensive care have their RR monitored continuously, but this equipment is rarely available on general hospital wards.
Objective: The primary objective is to assess the accuracy of the RespiraSense device in comparison with other methods currently used in clinical practice. The secondary objective is to assess the accuracy of the RespiraSense device in participants in different positions and when reading aloud.
Methods: A single-center, prospective observational study will investigate the agreement of the RespiraSense device as compared with other device measurements (capnography, electrocardiogram) and the current standard measurement of RR (manual counting by a trained health care professional). The different methods will be employed concurrently on the same participant as part of a single study visit.
Results: Recruitment to this study has not yet started as funding decisions are still pending. Therefore, results are not available at this stage. It is anticipated that the data required could be collected within 2 months of first recruitment to the study and data analysis completed within 6 months of the study start date.
Conclusions: The Evaluation of Agreement of Breathing Rates Measured by a Novel Device, Manual Counting, and Other Techniques Used in Clinical Practice (VENTILATE) study will provide further validation of the use of the RespiraSense device in subjects with abnormal respiratory rates.
| Reference Key |
chauhan2020jmiran
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | Anoop J Chauhan; |
| Journal | jmir research protocols |
| Year | 2020 |
| DOI |
doi:10.2196/15437
|
| URL | |
| Keywords | Keywords not found |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.