determination of uric acid in human serum by an enzymatic method using n-methyl-n-(4-aminophenyl)-3-methoxyaniline reagent
Clicks: 245
ID: 258861
2003
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
Popular Article
30.0
/100
245 views
39 readers
AI Quality Assessment
Not analyzed
Readership in this journal
PopularRanked #75 of 151 articles by views in meditsinskaia radiologiia
Most read
Least read
Bar heights use a square-root scale. Only the 120 most-read articles are drawn; the journal has 151 in total.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
In this work a new enzymatic method for the determination of uric acid in human serum has been developed. The method is based on the oxidative coupling reaction between the N-methyl-N-(4-aminophenyl)-3-methoxyaniline (NCP) reagent and the hydrogen donor reagent, N-ethyl-N-(2-hydroxy-3-sulfopropyl)-3-methylaniline (TOOS), in the system involving three enzymes: uricase, peroxidase and ascorbate oxidase. Using this method uric acid could be determined in concentrations up to 1.428 mmol/L, with a relative standard deviation of up to 1.8 %. The effect of the medium pH and the NCP concentration on the linearity of the chromogen absorbance versus the uric acid concentration curve was investigated. The influence of the uricase activity on the maximum rate of uric acid oxidation was also examined. The use of the NCP reagent demonstrated a more precise and more sensitive determination of the uric acid compared to the determination with 4-aminoantipyrine (4-AA) as the coupling regent. The sensitivity of the method determined from the calibration curve was 0.71 absorbance units per mmol/L of uric acid; the limit of detection was LOD = 0.0035 mmol/L and the limit of quantification was LOQ = 0.015 mmol/L of uric acid.
| Reference Key |
stankov2003journaldetermination
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | ;DEJAN STANKOV;PREDRAG DJURDJEVIC;MILENA JELIKIC-STANKOV |
| Journal | meditsinskaia radiologiia |
| Year | 2003 |
| DOI |
DOI not found
|
| URL | |
| Keywords |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.