magnetic solvent bar liquid-phase microextraction followed by gas chromatography-flame ionization detection for the trace determination of selected polycyclic aromatic hydrocarbons in environmental water samples
Clicks: 166
ID: 243204
2018
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
Emerging Content
30.0
/100
166 views
26 readers
AI Quality Assessment
Not analyzed
Readership in this journal
EmergingRanked #78 of 91 articles by views in ege eğitim dergisi
Most read
Least read
Bar heights use a square-root scale.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
A novel and efficient hollow fiber-based method, viz. magnetic solvent bar liquid-phase microextraction (MSB-LPME) combined with gas chromatography-flame ionization detection (GC-FID) was successfully developed for the trace determination of selected polycyclic aromatic hydrocarbons (PAHs) in environmental water samples. The target analytes were extracted from sample solution to the organic solvent immobilized in a fiber. After extraction, the analyte-adsorbed magnetic solvent bar could be readily isolated from the sample solution by a magnet which could greatly simplify the operation and also decline the total pretreatment time. The bar was first eluted with methanol, evaporated to dryness while the residue was dissolved in toluene and finally injected into GC-FID. Begin with, effective parameters controlling the performance of the microextraction were evaluated and optimized. The values of the detection limit of the method were in the range of 0.05-0.08 µg L-1 and the RSD% values for the analysis of 10.0 µg L-1 of the analytes was below than 5.8% (n= 6). A good linearity (0.998 ≥ r2 ≥ 0.994) and a broad linear range (0.1-200 µg L-1) were obtained. The method was eventually utilized for the preconcentration and determination of the PAHs in environmental water samples and satisfactory results were obtained.
| Reference Key |
farahani2018journalmagnetic
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | ;Hadi Farahani;Mohsen Zeeb |
| Journal | ege eğitim dergisi |
| Year | 2018 |
| DOI |
DOI not found
|
| URL | |
| Keywords |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.