evaluation of dose distribution accuracy in hdr brachytherapy of esophagus cancer based on mri normoxic polymer gel dosimetry

Clicks: 221
ID: 206222
2010
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Popular

Ranked #50 of 81 articles by views in the international archives of the photogrammetry, remote sensing and spatial information sciences

Most read Least read

Bar heights use a square-root scale.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
Introduction: The purpose of this work was to study the ability of MRI normoxic polymer gel dosimetry for evaluating the dose distribution in HDR brachytherapy of esophagial cancer at Imam Reza brachytherapy center (Mashhad, Iran). Materials and Methods: Initially, 2liters of normoxic gel (MAGIC) was fabricated and then poured into 12 calibration test tubes and placed in a perspex walled phantom. The gel phantom was irradiated with a brachytherapy remote-afterloader unit using a cobalt-60 brachytherapy source and the test tubes were irradiated with a range of known doses with a cobalt-60 teletherapy unit. Imaging was performed with a multi-spin-echo protocol and a T2 quantitative technique using a Siemens 1.5 T MRI machine. The MRI images were transferred to a computer and then image processing was performed in the MATLAB environment to extract R2 maps of the irradiated area. Results: In this study and at the reference point, the dose deviation between the gel dosimetry and the calculated data was 4.5%. The distance to agreement (DTA) for dose profiles was 2.7 mm. Also, dose sensitivity of the MAGIC gel dosimeter was 0.693 S-1Gy-1 (R2 =0.9376). Conclusion: In this work, the data obtained from TPS calculations were found in very good agreement with the measured results provided by gel dosimetry. It was evaluated using a comparison of isodoses and dose at the reference point, and dose profile verification. It is also concluded that the gel dosimetry systems have proven to be a useful tool for dosimetry in clinical radiotherapy applications.
Reference Key
toossi2010iranianevaluation Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors ;Mohammad Taghi Bahreyni Toossi;Ghorban Safaeian;Mohammad Hosein Bahreyni Toossi;Shahram Baiani
Journal the international archives of the photogrammetry, remote sensing and spatial information sciences
Year 2010
DOI
10.22038/ijmp.2010.7267
URL
Keywords

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.