in vivo and in vitro evaluations of repeatability and accuracy of vita easyshade® advance 4.0 dental shade-matching device
Клики: 479
ID: 197084
2015
Метрики качества и эффективности статьи
Общее качество
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Вовлечённость читателей
Popular Article
30.0
/100
479 просмотры
50 читатели
Оценка качества ИИ
Не проанализировано
Readership in this journal
PopularRanked #1 of 49 articles by views in Cognition
Most read
Least read
Bar heights use a square-root scale.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Аннотация
Objectives: The objective of this study was to evaluate the intra-device repeatability and accuracy
of dental shade-matching device (VITA Easyshade® Advance 4.0) using both in vitro and in vivo
models. Materials and methods: For the repeatability assessment, the in vivo model utilized
shade-matching device to measure the central region of the labial surface of right maxillary central
incisors of 10 people twice. The following tooth colors were measured: B1, A1, A2, A3, C1
and C3. The in vitro model included the same six Vitapan Classical tabs. Two measurements were
made of the central region of each shade tab. For the accuracy assessment, each shade tab from
3 Vitapan Classical shade guides was measured once. CIE L*a*b* values were determined. Intraclass
correlation coefficients (ICCs) were used to analyze the in vitro and in vivo intra-device repeatability
of the shade-matching device. The difference between in vitro and in vivo models was
analyzed. Accuracy of the device tested was calculated. Results: The mean color differences for
in vivo and in vitro models were 3.51 and 1.25 E units, respectively. The device repeatability ICCs
for in vivo measurements ranged from 0.858 to 0.971 and for in vitro from 0.992 to 0.994. Accuracy
of the device tested was 93.75%. Conclusion: Within the limitations of the experiment, VITA
Easyshade®Advance 4.0 dental shade-matching device enabled reliable and accurate measurement.
It can be a valuable tool for the determination of tooth colours.
| Ссылочный ключ |
ille2015actain
Используйте этот ключ для автоцитирования в рукописи при использовании
SciMatic Manuscript Manager или Thesis Manager
|
|---|---|
| Авторы | ;Davor Illeš;Iva Ž Alajbeg;Maja Žagar;Dubravka KNEZOVIĆ ZLATARIĆ |
| Журнал | Cognition |
| Год | 2015 |
| DOI |
DOI не найден
|
| URL | |
| Ключевые слова |
Цитирования
Цитирования не найдены. Чтобы добавить цитирование, свяжитесь с администратором по адресу info@scimatic.org
Комментарии
Комментариев пока нет. Будьте первым, кто прокомментирует эту статью.