validation of three viable-cell counting methods: manual, semi-automated, and automated
Clicks: 117
ID: 193569
2015
Article Quality & Performance Metrics
Overall Quality
Improving Quality
0.0
/100
Combines engagement data with AI-assessed academic quality
Reader Engagement
Emerging Content
2.1
/100
7 views
7 readers
Trending
AI Quality Assessment
Not analyzed
Abstract
A viable cell count is essential to evaluate the kinetics of cell growth. Since the hemocytometer was first used for counting blood cells, several variants of the methodology have been developed towards reducing the time of analysis and improving accuracy through automation of both sample preparation and counting. The successful implementation of automated techniques relies in the adjustment of cell staining, image display parameters and cell morphology to obtain equivalent precision, accuracy and linearity with respect to the hemocytometer. In this study we conducted the validation of three trypan blue exclusion-based methods: manual, semi-automated, and fully automated; which were used for the estimation of density and viability of cells employed for the biosynthesis and bioassays of recombinant proteins. Our results showed that the evaluated attributes remained within the same range for the automated methods with respect to the manual, providing an efficient alternative for analyzing a huge number of samples.
| Reference Key |
cadena-herrera2015biotechnologyvalidation
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | ;Daniela Cadena-Herrera;Joshua E. Esparza-De Lara;Nancy D. Ramírez-Ibañez;Carlos A. López-Morales;Néstor O. Pérez;Luis F. Flores-Ortiz;Emilio Medina-Rivero |
| Journal | PLoS genetics |
| Year | 2015 |
| DOI |
10.1016/j.btre.2015.04.004
|
| URL | |
| Keywords |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.