flow characteristics of the medtronic corevalve: difficulties estimating aortic valve cross-sectional area following transcatheter aortic valve implantation
Clicks: 165
ID: 185980
2014
Article Quality & Performance Metrics
Overall Quality
Improving Quality
0.0
/100
Combines engagement data with AI-assessed academic quality
Reader Engagement
Steady Performance
30.0
/100
164 views
12 readers
Trending
AI Quality Assessment
Not analyzed
Abstract
Background: Echocardiographic evaluation after transcatheter aortic valve implantation (TAVI) includes estimation of effective orifice area (EOA). EOA calculation depends on sub-valvular stroke volume (SV), which depends on sub-valvular diameter and velocity time integral (VTI). The Medtronic CoreValve area changes throughout its length. We aimed to (i) compare SV at two sites of flow acceleration: ‘pre-stent’ and ‘in-stent, pre-valve’, (ii) assess effects of possible differences in sub-valvular SV on EOA, and (iii) assess agreement of measurement of EOA calculation after CoreValve TAVI.
Methods: We studied 43 patients after CoreValve implantation. All had transthoracic echocardiography 5-7 days after TAVI. Sub-valvular SV was measured ‘pre-stent’ and ‘in-stent, pre-valve’. Measurement agreement was assessed by root mean square (RMS) differences and Bland-Altman analyses.
Results: SV was consistently higher ‘in-stent, pre-valve’ compared with ‘pre-stent’ (62±20ml vs. 53±19ml, p<0.001), so that EOA was correspondingly larger using ‘in-stent, pre-valve’ measurements (1.7±0.5cm2 vs. 1.4±0.5cm2, p<0.001). Between-observer RMS difference for calculation of EOA was higher ‘in-stent, pre-valve’ compared to ‘pre-stent’ (0.53 cm2 vs. 0.23cm2, difference from zero 0.17, p=0.002). Though sub-valvular diameter measurements were variable, VTI variability was additionally higher ‘in-stent, pre-valve’ compared to ‘pre-stent’ (0.42cm vs. 0.6cm, difference from zero -1.74, p=0.11).
Conclusion: Calculation of EOA after CoreValve TAVI is highly dependent on sub-valvular sample position. EOA may be underestimated using ‘pre-stent’ SV, and overestimated using ‘in-stent, pre-valve’ SV. Limitations in SV reproducibility suggests EOA should be used in conjunction with other indices of valve function in serial assessment of CoreValve function following TAVI.
Abstract Quality Issue:
This abstract appears to be incomplete or contains metadata (131 words).
Try re-searching for a better abstract.
| Reference Key |
duncan2014internationalflow
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | ;Alison Duncan;Eric Lim;Sarah Barker;Carlo Di Mario;Simon Davies;Neil Moat |
| Journal | transplant infectious disease : an official journal of the transplantation society |
| Year | 2014 |
| DOI |
https://doi.org/10.17987/icfj.v1i3.38
|
| URL | |
| Keywords |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.