testicular shield for para-aortic radiotherapy and estimation of gonad doses

Clicks: 58
ID: 166465
2008
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Popular

Ranked #26 of 28 articles by views in arya atherosclerosis

Most read Least read

Bar heights use a square-root scale.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
For radiotherapy of para-aortic and abdominal regions in male patients, gonads are to be protected to receive less than 2% of the prescribed dose. A testicular shield was fabricated for abdominal radiotherapy with 15 MV X-rays ((Clinac 2300 CD, Varian AG) with low melting point alloy (Cerroband). The dimensions of the testicular shield were 6.5 cm diameter and 3.5 cm depth with 1.5 cm wall thickness. During treatment, this shield was held in position by a rectangular sponge and Styrofoam support. Phantom measurement was carried out with a humanoid phantom and a 0.6 cc ion chamber. The mean energy of the scattered photon was calculated for single scattering at selected distances from the beam edge and with different field dimensions. One patient received radiotherapy with an inverted Y field and gonad doses were estimated using calibrated thermo-luminescent detector (TLD) chips. Measured doses with the ion chamber were 7.1 and 3.5% of the mid-plane doses without a shield at 3 and 7.5 cm off-field respectively. These values decreased to 4.6 and 1.7% with the bottom shield alone, and to 1.7 and 0.8% with both bottom and top shields covering the ion chamber. The measured doses at the gonads during the patient′s treatment were 0.5-0.92% for the AP field (0.74 ± 0.17%, n = 5) and 0.5-1.2% for the PA field (0.88 ± 0.24%, n = 5). The dose received by the testis for the full course of treatment was 32 cGy (0.8%) for a total mid-plane dose of 40 Gy. The first-scatter energy estimated at the gonads is around 1.14 MeV for a primary beam of 15 MV for a long axis dimension of 37 cm of primary field. During the patient′s treatment, the estimated absorbed doses at the gonads were comparable with reported values in similar treatments. The testicular shield reported in this study is of light weight and could be used conveniently in treatments of abdominal fields.
Reference Key
r2008journaltesticular Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors ;Ravichandran R;Binukumar J;Kannadhasan S;Shariff M;Ghamrawy Kamal
Journal arya atherosclerosis
Year 2008