optimisation of the low-frequency response of a multi-stage bioelectric amplifier for electrocardiogram (ecg) recording

Clicks: 228
ID: 163354
2016
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Steady

Ranked #50 of 435 articles by views in acta botânica brasílica

Most read Least read

Bar heights use a square-root scale. Only the 120 most-read articles are drawn; the journal has 435 in total.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
This paper reports the optimisation of the low-frequency response of a multi-stage bioelectric amplifier intended for use in the measurement of the human electrocardiogram (ECG) using high-impedance un-gelled electrodes. The frequency response was optimised to meet international performance requirements for electrocardiographic recorders, in particular the International Electrotechnical Commission 60601 standards [1,2]. The pole and zero locations of a multi-stage amplifier configuration were optimised to meet both time and frequency domain specifications. The optimum assignment for a three stage amplifier, having two differential stages and a differential-to-single-ended conversion stage was established by evaluating the performance of a number of circuit configurations. The optimum configuration was found to be two differential stages with a gain of 20dB each and a differential-to single-ended output stage with unity gain. The -3dB pole is placed at 0.028Hz and a zero at 0.0028Hz in the first and second stages to give an overall -3dB bandwidth of 0.043Hz. In addition, the pole of the input ac coupling network was placed at 0.0028Hz in order to meet the undershoot and recovery slope requirements in the narrow pulse response.
Reference Key
j.2016matecoptimisation Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors ;Burke Martin J.;Kofler Kerstin
Journal acta botânica brasílica
Year 2016
DOI
10.1051/matecconf/20167601008
URL
Keywords

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.