pacemaker failure due to loss of fluid seal in a patient with a 5/6 mm pacemaker header port and a 5 mm unipolar lead

Clicks: 109
ID: 152226
2013
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Steady

Ranked #52 of 63 articles by views in Medical decision making : an international journal of the Society for Medical Decision Making

Most read Least read

Bar heights use a square-root scale.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
Introduction: Before IS-1 (3.2 mm) standardization of pacemaker leads and connectors, 5/6 mm connector ports accomodated 5 mm or 6 mm diameter lead connector pins. Case report: A patient with sick sinus syndrome underwent implantation of a 5 mm unipolar atrial lead, mated to a 5/6 mm connector port Medtronic Spectrax Sx 5985 pacemaker. Pulse generator reached ERI in 2006, with change out to a Medtronic Sigma SSR306 (5/6 mm connector port) and preservation of the 5 mm lead. She was admitted in 2010 for atrial lead non capture from blood leak and corrosion of the header-connector pin apparatus. Discussion: 5/6 mm pacemaker header ports have a 5 mm flexible sealing ring at the port entrance to seal 5 mm or 6 mm lead connector pins. The inner barrel diameter of the connector port is 6 mm and insertion of a 5 mm lead results in a 0.5 mm tolerance circumferentially. Should the seal be compromised, blood can corrode the apparatus. To minimize this, we can employ (a) a cinching tie to further seal the silicone ring (b) universal adaptor sleeves (c) splice kits (d) lead adaptor kits. Aging leads, adaptor kits or sleeves themselves can result in lead failure. It may be safer to re-implant the entire system. Conclusion: A 5/6 mm configuration pacemaker header connector port allows for significant tolerances when a 5 mm lead is used. Consideration must be made to prevent leaks.
Reference Key
mbbs2013indianpacemaker Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors ;Paul Chun Yih Lim, MBBS;Boon Yew Tan;Kah Leng Ho;Chee Keong Ching;Wee Siong Teo
Journal Medical decision making : an international journal of the Society for Medical Decision Making
Year 2013
DOI
10.1016/S0972-6292(16)30691-X
URL
Keywords

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.