Attenuation correction of a flat table top for radiation therapy in hybrid PET/MR using CT- and Ge/Ga transmission scan-based μ-maps.
Clicks: 346
ID: 12989
2019
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
Steady Performance
68.4
/100
346 views
255 readers
Trending
AI Quality Assessment
Not analyzed
Readership in this journal
SteadyRanked #11 of 25 articles by views in Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)
Most read
Least read
Bar heights use a square-root scale.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
Hybrid PET/MR offers new opportunities in radiation oncology for tissue/tumour characterisation and response assessment. Attenuation correction (AC) is an important issue especially in the presence of immobilization devices and flat table tops (FTT). The goal of this study was to compare two methods of AC using CT- and Ge/Ga transmission scan-based attenuation maps (μ-maps) for a custom-designed FTT. Measurements were performed in the mMR PET/MR and TrueV PET/CT Biograph Siemens scanners with three different phantoms, namely the Siemens MR-QA, a cubic canister and the NEMA IEC body phantom. The study revealed that the MR image quality is not hampered by the presence of the FTT. For cubic canister applying the scanner's inherent AC alone resulted in inaccuracies in PET images, with up to -4.0% underestimation of the activity. The mean NEMA sphere activity measurements without FTT, agreed within 3.5% with the respective inserted activity. Placing the FTT in the PET/MR scanner resulted in a difference to the injected activity of 4.5% when the table was not corrected for. By introducing the μ-maps the discrepancy between the used activity and the measurements decreased down to 2.6%. To improve the AC of the FTT the creation of a dedicated μ-map was necessary while the CT-based μ-map performed equally good as the source transmission scan-based one.
| Reference Key |
witoszynskyj2019attenuationphysica
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | Witoszynskyj, Stephan;Andrzejewski, Piotr;Georg, Dietmar;Hacker, Marcus;Nyholm, Tufve;Rausch, Ivo;Knäusl, Barbara; |
| Journal | Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) |
| Year | 2019 |
| DOI |
S1120-1797(19)30185-1
|
| URL | |
| Keywords | Keywords not found |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.