DEMONSTRATION OF THE POTENTIAL AND DIFFICULTIES OF COMBINED TL AND OSL MEASUREMENTS OF TLD-600 AND TLD-700 FOR THE DETERMINATION OF THE DOSE COMPONENTS IN COMPLEX NEUTRON-GAMMA RADIATION FIELDS.

Clicks: 234
ID: 97178
2020
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Steady

Ranked #20 of 54 articles by views in Radiation protection dosimetry

Most read Least read

Bar heights use a square-root scale.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
The results reported herein demonstrate the potential application of combined optically stimulated luminescence/thermoluminescent (OSL/TL) measurements in neutron-gamma discrimination dosimetry. The advantages of OSL/TL are two-fold: (i) The OSL and TL readout can be carried out on the same sample and (ii) the greater efficiency of OSL to high ionization density radiation due to F 2 and F3 excitation. The gamma/electron calibration coefficients for LiF:Mg, Ti (TLD-600 and TLD-700) were measured using a 90Sr/90Y source calibrated at the SARAF-SSDL nuclear facility. The estimation of the neutron calibration coefficients was carried out by irradiation with broad-spectrum beam of fast neutrons with median energy 5 MeV at the Radiological Research Accelerator Facility (RARAF) of Columbia University. Naturally cooled samples of TLD-600 and TLD-700 were dosed to levels of 29.8 Gy neutrons and 6.1 Gy gammas in air and KERMA calculations employed to transfer the levels of dose to6,7LiF. A figure of merit for fast-neutron/gamma ray discrimination was determined at 10.6 for TLD-700 in the current measurements. The use of combined TLD-600/TLD-700 allowed, as well, the determination of a considerable and somewhat unexpected thermal neutron component of 116 Gy in TLD-600.
Reference Key
oster2020demonstrationradiation Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors
Journal Radiation protection dosimetry
Year 2020
DOI
ncz297
URL
Keywords Keywords not found

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.