Robust blood pressure estimation from finger photoplethysmography using age-dependent linear models.

Clicks: 386
ID: 94920
2020
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Emerging

Ranked #5 of 16 articles by views in physiological measurement

Most read Least read

Bar heights use a square-root scale.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
This work aims to develop an efficient and robust age-dependent multiple linear regression (MLR) model to estimate blood pressure (BP) from a single-source photoplethysmography (PPG) and biometrics, which could be embedded in the microcontroller (MCU) of pulse oximeters.Hemodynamic features were extracted from the PPG signal using its waveform, derivatives, and biometrics. Whole-based, feature-based, and fusion models were evaluated and compared for different age groups. Their performance was tested using 1086 subjects with a leave-one-subject-out cross-validation. The improvement by adding biometrics and the long-term calibration effect were investigated in detail. The relative importance of each feature was compared between different age groups and the implication was discussed.Fusion model achieved the best performance in subjects with well-defined PPG features, whereas the feature-based method was better suited for subjects with damped signals. Adding age significantly improved both systolic BP (SBP) and diastolic BP (DBP) estimation accuracy for older subjects (>50 years old) with well-defined features, while it only improved diastolic BP accuracy for older subjects with damped signals. For younger subjects (≤50 years old), the contribution of age was very small. A simple subtraction of subject-specific calibration factors significantly reduced biometric related errors, which also improved the linearity of BP estimation. The relative importance analysis of input features suggests that separate models are indeed necessary for different age groups with different signal qualities, especially for DBP estimation in older subjects.This study shows a reasonable BP estimation accuracy with age-dependent MLR models, which may help to equip current pulse oximeters with additional functionalities.
Reference Key
xing2020robustphysiological Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors Xing, Xiaoman;Ma, Zhimin;Zhang, Mingyou;Gao, Xi;Li, Ying;Song, Mingxuan;Dong, Wen-Fei;
Journal physiological measurement
Year 2020
DOI
10.1088/1361-6579/ab755d
URL
Keywords

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.