A low oxygen consumption pneumatic ventilator for emergency construction during a respiratory failure pandemic.
Clicks: 498
ID: 77993
2010
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
Popular Article
30.0
/100
498 views
51 readers
AI Quality Assessment
Not analyzed
Readership in this journal
PopularRanked #1 of 23 articles by views in Anaesthesia
Most read
Least read
Bar heights use a square-root scale.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
The UK influenza pandemic plan predicts up to 750,000 additional deaths with hospitals prioritising patients against inadequate resources. We investigated three prototype low-cost, gas-efficient, pneumatic ventilators in a test lung model at different compliance and rate settings. Mean (SD) oxygen consumption was 0.913 (0.198) and 1.119 (0.267) l.min(-1) at tidal volumes of 500 ml and 700 ml respectively. Values of F(I)o(2) increased marginally as lung compliance reduced, reflecting the increased ventilator workload and consequent increased enrichment of breathing gas by waste oxygen from the pneumatic mechanism. We also demonstrated that a stable nitric oxide concentration could be delivered by this design following volumetric principles. It is possible to make a gas-efficient ventilator costing less than 200 pounds from industrial components for use where oxygen is available at 2-4 bar, with no pressurised air or electrical requirements. Such a device could be mass-produced for crises characterised by an overwhelming demand for mechanical ventilation and a limited oxygen supply.
| Reference Key |
williams2010aanaesthesia
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | Williams, D;Flory, S;King, R;Thornton, M;Dingley, J; |
| Journal | Anaesthesia |
| Year | 2010 |
| DOI |
10.1111/j.1365-2044.2009.06207.x
|
| URL | |
| Keywords |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.