Optimized electrical bioimpedance measurements of abdominal wall on a porcine model for the continuous non-invasive assessment of intra-abdominal pressure.

Clicks: 226
ID: 69225
2019
Article Quality & Performance Metrics
Overall Quality Improving Quality
0.0 /100
Combines engagement data with AI-assessed academic quality
AI Quality Assessment
Not analyzed
Abstract
This work describes the optimization of electrical bioimpedance measurements for indirect intra-abdominal pressure (IAP) assessment. The experimental run was performed on a female Sus scrofa domesticus (domestic pig). Different values of IAP were induced by inflation of the abdominal cavity, using a trocar placed near the umbilicus over the linea alba. The whole experiment was run within 1 h of the subject being sacrificed. The abdominal wall thickness was measured at an IAP of 5 mmHg. An exponential trend linking between the bioimpedance values at 99.8 kHz and the IAP was found. Non-optimized electrode placement presented a strongly reduced sensitivity to IAP changes above 7 mmHg. Upon optimization and placing the electrodes with a separation of about 3.6 times the measured abdominal wall thickness, the sensitivity for high IAP drastically increased, allowing continuous non-invasive monitoring of IAP, confirming the optimization method proposed in this work.
Reference Key
david2019optimizedjournal Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors David, Marcelo;Amran, Omer;Peretz, Aviad;Raviv, Aviad;Pracca, Francisco;
Journal Journal of clinical monitoring and computing
Year 2019
DOI
10.1007/s10877-019-00441-6
URL
Keywords

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.