A high-precision, geometric and registration accuracy full-system test method for adaptive SRS, demonstrated on Gamma Knife® Icon™.

Clicks: 333
ID: 63109
2019
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
A novel full-system test (FST) phantom and method have been developed to demonstrate and quality assure the geometric accuracy of image co-registration and overall shot delivery in the context of SRS using Gamma Knife® Icon™. The method uses Vernier scale bars to achieve sub-voxel precision co-registration measurements and pin-located radiochromic films to determine overall shot delivery precision. Validation tests demonstrated that artificially applied registration errors of < 0.15 mm could be accurately detected and quantified. Cross-validation of full-system test results with the manufacturer standard focal precision test demonstrated that both approaches measure similar focal precision errors, to within < 0.1 mm, and that registration and focal precision components of the full-system geometric error can be successfully decoupled using our Vernier FST approach. CBCT co-registration errors were shown to be of comparable magnitude to the focal precision errors, demonstrating that CBCT registration based in-mask treatments can achieve sub-voxel geometric accuracy, rivalling traditional frame-based immobilisation. This full-system geometric test method and phantom design concept is in principle applicable to any SRS technique involving image co-registration.
Reference Key
nix2019ajournal Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors
Journal journal of radiosurgery and sbrt
Year 2019
DOI
DOI not found
URL URL not found
Keywords

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.