Effect of Material and Wall Thickness Buildup Caps on the Head Scatter Factor Measurements in Irregular Fields Shielded by Cerrobend
Clicks: 391
ID: 51360
2017
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
Steady Performance
66.9
/100
391 views
294 readers
Trending
AI Quality Assessment
Not analyzed
Readership in this journal
SteadyRanked #1 of 6 articles by views in journal of medical signals and sensors
Most read
Least read
Bar heights use a square-root scale.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
The head scatter factor (Sc) is important to measurements radiation beam and beam modeling of treatment planning systems used for advanced radiation therapy techniques. This study aimed to investigate the design of a miniphantom to measurement variations in collimator Sc in the presence of shielding blocks for shaping the beam using different field sizes. Copper, Brass, and Perspex buildup caps were designed and fabricated locally as material with three different thicknesses for buildup caps (miniphantoms). Measurements were performed on an Elekta Compact medical linear accelerator (6 MV) in Shafa Kerman Hospital, Iran. The Farmer-type ion chamber FG65-P (Scanditronix, Wellhofer) was used for all measurements. To measure the Sc, miniphantom was positioned in a stand vertical to the beam central axis. The data indicate that the Sc measurements using different buildup cap materials and thicknesses in 5 × 10, 7.5 × 7.5, and two 10 × 10 cm Cerrobend shield blocks ranged 0.98 to 1.00, 1.04 to 1.05, and 1.04 to 1.06, respectively. Also, it was observed that by increasing the block shield area from 50 cm2 to both 56.25 and 100 cm2, the Sc increased in all situations. Results showed that using Brass compared to Perspex and Copper has less uncertainty due to its simple preparation and cutting which is useful to measurement of variations in collimator Sc and shaping the photon beam.
| Reference Key |
jomehzadeh2017effectjournal
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | Jomehzadeh, Ali;Shahbazi-Gahrouei, Daryoush;Jahanbakhsh, Vahid;, ; |
| Journal | journal of medical signals and sensors |
| Year | 2017 |
| DOI |
DOI not found
|
| URL | |
| Keywords |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.