Influence of physical assessment of different light-curing units on irradiance and composite microhardness top/bottom ratio.

Clicks: 420
ID: 46565
2016
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Popular

Ranked #1 of 4 articles by views in odontology

Most read Least read

Bar heights use a square-root scale.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
The aim of this study was to evaluate the influence of the physical assessment of different light-curing units from 55 dental offices on the irradiance and composite microhardness top/bottom ratio, and the influence of the radiometers for LED or QTH light sources on irradiance measurement. The irradiance of each light-curing unit was evaluated with two radiometers, either for LED or QTH light. A questionnaire regarding the type of source (LED or QTH), time of use, date of last maintenance and light-curing performance assessment applied. The physical assessments were evaluated regarding damage or debris on the light tip. For each light-curing unit, three composite specimens were made (diameter = 7 mm; thickness = 2 mm) with polymerizing time of 20 s, in order to perform the microhardness (Knoop) test. Data were analyzed by Kruskal-Wallis and Dunn test (α = 0.01). There was wide variation in irradiance (0-1000 mW/cm(2)). Approximately 50 % of the light-curing units presented radiation lower than 300 mW/cm(2); 10 % of light-curing units, especially those with LED source, presented values higher than 800 mW/cm(2), and 43 % of light-curing units worked with adequate irradiance between 301 and 800 mW/cm(2). In almost 60 % of cases, no maintenance of light-curing units was performed in a period of 3 to 10 years. The age of the light-curing units and the use of inadequate tips interfered negatively in irradiance. The data emphasize the importance of periodic maintenance of light-polymerizing, light-curing units.
Reference Key
morimoto2016influenceodontology Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors Morimoto, Susana;Zanini, Renata Aló Maluza;Meira, Josete Barbosa Cruz;Agra, Carlos Martins;Calheiros, Fernanda Calabró;Nagase, Denis Yudi;
Journal odontology
Year 2016
DOI
10.1007/s10266-015-0229-y
URL
Keywords

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.