A Starling-Like Total Work Controller for Rotary Blood Pumps: An In Vitro Evaluation.

Clicks: 260
ID: 43767
2019
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Steady

Ranked #6 of 8 articles by views in artificial organs

Most read Least read

Bar heights use a square-root scale.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
Due to improved durability and survival rates, rotary blood pumps (RBPs) are the preferred left ventricular assist device when compared to volume displacement pumps. However, when operated at constant speed, RBPs lack a volume balancing mechanism which may result in left ventricular suction and sub-optimal ventricular unloading. Starling-like controllers have previously been developed to balance circulatory volumes, however they do not consider ventricular workload as a feedback and may have limited sensitivity to adjust RBP workload when ventricular function deteriorates or improves. To address this, we aimed to develop a Starling-like total work controller (SL-TWC) that matched the energy output of a healthy heart by adjusting RBP hydraulic work based on measured left ventricular stroke work and ventricular preload. In a mock circulatory loop, the SL-TWC was evaluated using a HeartWare HVAD in a range of simulated patient conditions. These conditions included changes in systemic hyper- and hypo-tension, pulmonary hypertension, blood circulatory volume, exercise and improvement and deterioration of ventricular function by increasing and decreasing ventricular contractility. The SL-TWC was compared to constant speed control where RBP speed was set to restore cardiac output to 5.0 L min at rest. Left ventricular suction occurred with constant speed control during pulmonary hypertension but was prevented with the SL-TWC. During simulated exercise, the SL-TWC demonstrated reduced LVSW (0.51 J) and greater RBP flow (9.2 L min ) compared to constant speed control (LVSW: 0.74 J and RBP flow: 6.4 L min ). In instances of increased ventricular contractility, the SL-TWC reduced RBP hydraulic work, whilst maintaining cardiac output similar to the rest condition. In comparison, constant speed overworked and increased cardiac output. The SL-TWC balanced circulatory volumes by mimicking the Starling mechanism, whilst also considering changes in ventricular workload. Compared to constant speed control, the SL-TWC may reduce complications associated with volume imbalances, adapt to changes in ventricular function and improve patient quality of life.
Reference Key
wu2019aartificial Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors Wu, Eric;Stevens, Michael;Nestler, Frank;Pauls, Jo;Bradley, Andrew;Fraser, John;Tansley, Geoff;Gregory, Shaun;
Journal artificial organs
Year 2019
DOI
10.1111/aor.13570
URL
Keywords

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.