UPLC-MS/MS method for the simultaneous determination of imatinib, voriconazole and their metabolites concentrations in rat plasma.
Clicks: 476
ID: 3448
2019
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
Emerging Content
78.8
/100
476 views
346 readers
Trending
AI Quality Assessment
Not analyzed
Readership in this journal
EmergingRanked #2 of 235 articles by views in Journal of pharmaceutical and biomedical analysis
Most read
Least read
Bar heights use a square-root scale. Only the 120 most-read articles are drawn; the journal has 235 in total.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
In the present study, a simple ultra-performance liquid chromatography tandem mass spectrometry (UPLC-MS/MS) method used to measure the plasma concentrations of imatinib, voriconazole and their metabolites (N-desmethyl imatinib and N-oxide voriconazole) in rats simultaneously making use of diazepam as the internal standard (IS) had been developed and validated. A simple protein precipitation by acetonitrile was employed for the sample preparation, then the analytes (imatinib, voriconazole and their metabolites) were eluted on an Acquity UPLC BEH C18 column (2.1 mm × 50 mm, 1.7 μm) using the mobile phase that made up by acetonitrile (A) and 0.1% formic acid in water (B). In positive ion mode, four analytes and IS were monitored by multiple reaction monitoring (MRM) as the following mass transition pairs: m/z 494.3→394.2 for imatinib, m/z 480.3→394.2 for N-desmethyl imatinib, m/z 350.1→281.1 for voriconazole, m/z 366.1→224.1 for N-oxide voriconazole, and m/z 285.0→154.0 for IS. This method exhibited a good linearity for each analyte. Inter-day and intra-day precision were determined with values of 0.3-14.8% and 2.6-14.8%, respectively; the accuracy values were from -12.5% to 10.2%. Finally, data of matrix effect, extraction recovery, and stability were all conformed to the bioanalytical method validation of acceptance criteria of FDA recommendations. This method is an efficient tool for simultaneous determination of the four analytes and has been successfully applied for pharmacokinetic study in rats.
| Reference Key |
xu2019uplcmsmsjournal
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | Xu, Ren-Ai;Lin, Qianmeng;Qiu, Xiangjun;Chen, Jingjing;Shao, Yuanyuan;Hu, Guoxin;Lin, Guanyang; |
| Journal | Journal of pharmaceutical and biomedical analysis |
| Year | 2019 |
| DOI |
S0731-7085(18)32461-0
|
| URL | |
| Keywords | Keywords not found |