Balloon shaft fracture, deflation failure, and intracoronary entrapment: A stepwise percutaneous bailout – A case Report

Clicks: 7
ID: 337471
2026
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Emerging

Ranked #385 of 408 articles by views in European Heart Journal - Case Reports

Most read Least read

Bar heights use a square-root scale. Only the 120 most-read articles are drawn; the journal has 408 in total.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
Abstract Background Balloon catheter fracture during percutaneous coronary intervention (PCI) is rare but potentially catastrophic. The literature describes three distinct forms: shaft fracture, undeflatable balloon, and entrapped balloon. Published bailout strategies typically address each problem in isolation; however, no specific guidance exists when multiple forms occur simultaneously. Case summary A 62-year-old man with multivessel coronary disease underwent staged PCI of an a left circumflex–obtuse marginal bifurcation. During lesion preparation, a reused semi-compliant balloon (2.5 × 14 mm) fractured after multiple inflations, leaving an inflated, entrapped balloon fragment at the bifurcation with impaired antegrade flow. A stepwise percutaneous bailout was performed: (1) length comparison with an intact reference balloon suggested that a long fragment was partly retained inside the guiding catheter; (2) a guide extension catheter was advanced to displace the fragment distally, achieving passive deflation and restoring flow; and (3) the deflated fragment was retrieved using a 2.0 mm trapping balloon straddling the tip of the guiding catheter. The procedure was completed successfully with stenting, and the patient recovered uneventfully. Discussion Sequential management converted a high-risk situation into a tractable single problem. Distal displacement of the inflated fragment relieved ischaemia before any retrieval attempt was made. Trapping with a second balloon straddling the guide catheter tip may improve grip when the fragment partly protrudes from the guiding catheter.
Reference Key
openalex_W7223939440 Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors
Journal European Heart Journal - Case Reports
Year 2026
DOI
10.1093/ehjcr/ytag804
URL
Keywords Keywords not found

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.