Successful extravascular ICD implantation in a patient with ankylosing spondylitis and severe thoracic deformity: a case report

Clicks: 10
ID: 332495
2026
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Emerging

Ranked #360 of 407 articles by views in European Heart Journal - Case Reports

Most read Least read

Bar heights use a square-root scale. Only the 120 most-read articles are drawn; the journal has 407 in total.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
Abstract Background Severe thoracic deformity in ankylosing spondylitis (AS) can complicate ICD implantation and device selection. Substernal EV-ICD implantation requires careful anatomical assessment and individualized procedural planning. Case Summary A 57-year-old male with severe AS and dilated cardiomyopathy was referred for primary prevention ICD implantation. Given the high complication risks associated with conventional approaches, EV-ICD was selected. A tailored positioning strategy was employed to accommodate the thoracic deformity. Intraoperative resistance during substernal dissection was successfully overcome through meticulous re-dissection. Favourable intraoperative electrical parameters were obtained, including an R-wave sensing amplitude of 5.6 mV without P-wave oversensing and successful defibrillation at 30 J on the first attempt. At 6-month follow-up, device performance remained stable without sensing anomalies or inappropriate shocks. Discussion This case demonstrates the technical feasibility of EV-ICD implantation in a carefully selected patient with severe AS-associated thoracic deformity. Preoperative CT assessment, tailored patient positioning, careful dissection, and surgical backup enabled successful implantation and stable device performance during 6 months of follow-up.
Reference Key
openalex_W7222223659 Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors
Journal European Heart Journal - Case Reports
Year 2026
DOI
10.1093/ehjcr/ytag758
URL
Keywords Keywords not found

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.