LAA Occluder device impeding PFA ablation of an anterior mitral line: a case report about the role of a dual energy source

Clicks: 10
ID: 331557
2026
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Emerging

Ranked #344 of 408 articles by views in European Heart Journal - Case Reports

Most read Least read

Bar heights use a square-root scale. Only the 120 most-read articles are drawn; the journal has 408 in total.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
Abstract Background Pulsed field ablation and percutaneous closure of the left atrial appendage are increasingly used in patients with atrial fibrillation. Interaction between this form of energy and metallic structures in the heart are an emerging problem. We first report a LAAC limiting PFA using a novel dual energy ablation catheter, solved by switching for the classic radiofrequency ablation. Case summary An 85-year-old woman with HFpEF and a history of persistent atrial fibrillation (status post PVI in 2019, with a 25 mm Abbott Amulet LAAC implanted in 2023) presented with recurrent atrial tachycardia. During redo ablation, mapping revealed a macroreentrant circuit through an anterior left atrial low-voltage area. While extending a linear pulsed-field ablation (PFA) lesion toward the LAAC, the generator repeatedly aborted applications due to invalid impedance values near the device—despite the device having been in place for three years. Switching to radiofrequency energy on the same catheter allowed the line to be completed successfully, achieving bidirectional block and tachycardia non-inducibility. Discussion Interaction between PFA and metallic cardiac devices may limit catheter ablation, problem that can be solved by technologies that allow to switch for radiofrequency sources. Dual energy technologies may be therefore preferred in patients with LAAC devices.
Reference Key
openalex_W7220674405 Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors
Journal European Heart Journal - Case Reports
Year 2026
DOI
10.1093/ehjcr/ytag711
URL
Keywords Keywords not found

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.