Enantioselective multiple heartcut two-dimensional ultra-high-performance liquid chromatography method with a Coreshell chiral stationary phase in the second dimension for analysis of all proteinogenic amino acids in a single run.
Clicks: 360
ID: 32974
2018
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
Steady Performance
80.8
/100
360 views
241 readers
Trending
AI Quality Assessment
Not analyzed
Readership in this journal
SteadyRanked #11 of 58 articles by views in journal of chromatography a
Most read
Least read
Bar heights use a square-root scale.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
A multiple heartcut (MHC) 2D-UHPLC method with UV detection has been developed for the enantioselective analysis of complex amino acid mixtures in a single run. The MHC method is based on an achiral gradient RPLC separation with 1.8 μm C18 phase (100 × 2.1 mm ID column) in the first dimension (D) and enantioselective isocratic separation on a tert-butylcarbamoylquinine-based 2.7 μm Coreshell particle column (50 × 3 mm ID) in the second dimension (D). Pre-column derivatization has been performed with Sanger's reagent (2,4-dinitrofluorobenzene) yielding chromogenic 2,4-dinitrophenylated amino acids (DNP-AAs). Heartcuts of 40 μL fractions of the D peaks were sampled into the D system via a two-position four-port dual valve connected to two loop decks each equipped with six 40 μL parking loops. Using this setup, 25 amino acids (20 proteinogenic plus allo-Thr, allo-Ile, homoserine (Hse), Orn, β-Ala) have been analyzed enantioselectively in a fully automated manner with a single chiral column within 130 min total run time (D and D). All D separations together took 101.5 min (29 cuts with 3.5 min run time each) and thus the total analysis time was quite efficiently utilized. Faster separations were restricted by some software constraints which did not allow to adjust run times in D individually. The practical utility of this enantioselective MHC method is documented by application for the absolute configuration determination of the amino acids in gramicidin and bacitracin. Further optimizations should lead to a generic enantioselective amino acid analyzer for the quality control of synthetic peptides and the structural characterization of non-ribosomal peptides.
| Reference Key |
woiwode2018enantioselectivejournal
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | Woiwode, Ulrich;Neubauer, Stefan;Lindner, Wolfgang;Buckenmaier, Stephan;Lämmerhofer, Michael; |
| Journal | journal of chromatography a |
| Year | 2018 |
| DOI |
S0021-9673(18)30698-8
|
| URL | |
| Keywords | Keywords not found |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.