Lessons learned from Micra leadless pacemaker retrieval: a case series
Clicks: 5
ID: 321819
2026
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
0.0
/100
5 views
0 readers
AI Quality Assessment
Not analyzed
Readership in this journal
Ranked #3 of 247 articles by views in European Heart Journal - Case Reports
Most read
Least read
Bar heights use a square-root scale. Only the 120 most-read articles are drawn; the journal has 247 in total.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
Abstract Background Retrieving an implanted leadless pacemaker is a challenging component of its management. However, there are no dedicated tools specifically designed for retrieving the Micra leadless pacemaker. Given the lack of commercial retrieval systems, there is an urgent need to document and validate alternative interventional strategies to manage complications associated with device malfunction or migration. Case Summary In this series, we present cases involving increased capture threshold, device embolization into the right pulmonary artery, and broken tines following Micra implantation. We successfully completed four retrieval procedures using a coordinated approach with existing devices, including the Micra transcatheter pacing system, coronary angiography catheter, and endovascular snare system. Discussion Notably, these cases exhibited different characteristics during the retrieval of the Micra leadless pacemaker, which may assist colleagues in dealing with such scenarios. Our experience demonstrates that a strategy combining the Micra delivery catheter with an endovascular snare system, without the need for a steerable sheath, was sufficient and effective for successful Micra retrieval.
| Reference Key |
openalex_W7169868199
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | Qi Guo, Hanhua Gao, Zhihua Deng, Shanghai Li, Shuxian Zhou, Yi-Xiang Wang, Yangxin Chen |
| Journal | European Heart Journal - Case Reports |
| Year | 2026 |
| DOI |
10.1093/ehjcr/ytag539
|
| URL | |
| Keywords | Keywords not found |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.