Entrapment of a J-tip guidewire within the pulmonary vein during FARAPULSETM pulsed field ablation. A case report

Clicks: 11
ID: 320763
2026
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Steady

Ranked #3 of 283 articles by views in European Heart Journal - Case Reports

Most read Least read

Bar heights use a square-root scale. Only the 120 most-read articles are drawn; the journal has 283 in total.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
Abstract Background Pulsed field ablation (PFA) is a rapidly adopted technology for atrial fibrillation (AF) ablation and is associated with a favorable safety profile and a low risk of complications. However, mechanical complications have rarely been reported. Case summary A 37-year-old Asian male with paroxysmal AF underwent PFA using the FARAPULSETM system. Following successful ablation of the left superior pulmonary vein (PV), the J-tip guidewire advanced into the left inferior PV and became completely entrapped. Venography demonstrated that, although the guidewire tip remained intravascular, a proximal segment appeared to course along the vessel wall, suggesting passage through an intraluminal structure. Multiple attempts at withdrawal using push-pull maneuvers were unsuccessful. Repositioning with a radiofrequency ablation catheter also failed to release the entrapment. Given the risk of vessel rupture and major hemorrhage, the procedure was converted to general anesthesia in a hybrid operating room with surgical standby. Bridge occlusion balloon-assisted extraction successfully released the entrapped guidewire without vascular injury. The retrieved guidewire had adherent fibrous white tissue, suggesting entrapment by a membranous pulmonary venous structure. Discussion Balloon-assisted endovascular retrieval may represent a safe and less invasive alternative to surgical intervention for managing such entrapments. Operators should be aware of this rare but potentially hazardous complication.
Reference Key
openalex_W7168097484 Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors Ken Watanabe, Takanori Arimoto, Yuta Kobayashi, Naoaki Hashimoto, Masafumi Watanabe
Journal European Heart Journal - Case Reports
Year 2026
DOI
10.1093/ehjcr/ytag503
URL
Keywords Keywords not found

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.