Image sharpening improves the visibility of reference points on low-dose lateral cephalometric radiographs: a phantom study

Clicks: 39
ID: 319761
2026
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Steady

Ranked #40 of 59 articles by views in dentomaxillofacial radiology

Most read Least read

Bar heights use a square-root scale.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
This phantom study aimed to determine whether a novel non-artificial intelligence image-sharpening approach improves the visibility of reference points on low-dose lateral cephalometric radiographs. Lateral cephalometric radiographs were obtained using an adult human bone phantom at 100% of the routine dose. Low-dose imaging was performed using 60, 40, and 20% of the routine dose. Each low-dose image underwent a two-level image-sharpening process using the Medical Image Enhancer software. Six reviewers evaluated the visibility of the five reference points in the first and second sets of sharpened and 100% radiation-dose images. Using Scheffé's method of paired comparisons, odds ratios (OR) were reported as relative values for a standard cephalometric radiograph (100% radiation-dose) to compare the visibility of reference points among imaging conditions. For the first set of mildly sharpened 60%-dose images, the ORs for the four reference points were > 1.0, Point A (OR 1.12, p = 0.83), Nasion (OR 2.26, p = 0.11), posterior nasal spine (OR 2.35, p = 0.1), and incisal edge of the maxillary central incisor (OR 1.35, p = 0.53), indicating visibility comparable to or better than the 100%-dose images. For the second set of markedly sharpened images, rated visibility was significantly higher for all reference points at 60% and 40% doses and for several reference points at 20% dose. Image-sharpening algorithms improved the visibility of reference points on low-dose phantom radiographs, suggesting their technical feasibility as a post-acquisition approach for dose-reduced cephalometric imaging.
Reference Key
openalex_W7167498646 Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors
Journal dentomaxillofacial radiology
Year 2026
DOI
10.1093/dmfr/twag047
URL
Keywords Keywords not found

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.