Mechanical Interaction Between Sequential Micra Devices Resolved by Programming Alone: A Case of Premature Battery Depletion in a 91-Year-Old Patient
Clicks: 1
ID: 317444
2026
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
0.0
/100
1 views
0 readers
AI Quality Assessment
Not analyzed
Readership in this journal
Ranked #215 of 251 articles by views in European Heart Journal - Case Reports
Most read
Least read
Bar heights use a square-root scale. Only the 120 most-read articles are drawn; the journal has 251 in total.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
Abstract Background The Micra transcatheter pacing system provides a leadless alternative for patients requiring ventricular pacing. When a leadless device reaches end of service or fails prematurely, a second device is commonly implanted. Case presentation We report a 91-year-old man with premature Micra AV battery depletion 28 months after implantation. During the second implantation, mechanically mediated device–device interaction caused intermittent oversensing. Oversensing was resolved by increasing the sensing threshold, which avoided device repositioning or retrieval. Conclusion Suspected mechanically mediated interaction between sequential Micra devices may cause oversensing and pacing inhibition. In patients with clinically stable condition and adequate intrinsic R-wave amplitude, programming optimization may be considered before invasive revision; however, future management should account for cumulative right ventricular device burden.
| Reference Key |
openalex_W7164802904
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | 何昌见, Tianze Guo, Chunhua Ding, Wenchang Zhang |
| Journal | European Heart Journal - Case Reports |
| Year | 2026 |
| DOI |
10.1093/ehjcr/ytag431
|
| URL | |
| Keywords | Keywords not found |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.