Low-Dose Cone-Beam Computed Tomography in Implant Dentistry: A Clinical Validation of Diagnostic Sufficiency
Clicks: 1
ID: 314837
2026
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
0.0
/100
1 views
0 readers
AI Quality Assessment
Not analyzed
Readership in this journal
Ranked #26 of 46 articles by views in dentomaxillofacial radiology
Most read
Least read
Bar heights use a square-root scale.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
Abstract Aim To validate the diagnostic sufficiency of a low-dose cone-beam computed tomography (CBCT) protocol for dental implant assessment by comparing it to a standard-dose protocol in a clinical setting. The study evaluated linear measurement accuracy, subjective clarity of anatomical landmarks, and the impact on bone microarchitecture assessment. Materials and Methods This retrospective study analyzed pre-operative (standard-dose) and post-operative (low-dose) CBCT scans from 37 patients (37 standard-dose and 37 low-dose scans for both maxillary and mandibular sites). The low-dose protocol achieved a 76.7% dose reduction by decreasing tube current and exposure time. The diagnostic utility was assessed: 1) subjective clarity scoring of the mandibular canal and maxillary sinus floor; 2) objective linear measurements of key anatomical dimensions; and 3) quantitative analysis of bone microstructural parameters (Tb.N, Tb.Sp, Tb.Th, BV/TV). Reliability assessment was also performed for clarity scoring and quantitative measurements. Statistical comparisons were performed using Chi-square and paired-samples t-tests. Results No statistically significant differences were found between the low-dose and standard-dose protocols in the subjective clarity of anatomical structures (P > 0.05) or in the accuracy of linear measurements (P > 0.05). However, all four bone microstructural parameters showed statistically significant differences between the two protocols in both the maxilla and mandible (P < 0.05 for all). Intra- and inter-examiner assessments demonstrated substantial to good consistency across all qualitative (weighted Kappa >0.611) and quantitative (ICC >0.702) parameters. Conclusion This study provides clinical evidence that for the specific CBCT unit tested, the optimized low-dose CBCT protocol reduces patient radiation exposure by over 76.7% for dental implant assessment without compromising the diagnostic accuracy required for linear measurements and visualization of key anatomical landmarks. Therefore, for the tested model, the low-dose protocol is sufficient for pre-operative CBCT scanning in implant dentistry. This offers a clinically validated method for enhancing patient safety in this specific context, effectively applying the ALADAIP principle.
| Reference Key |
openalex_W7162295354
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | Ruiyang Ni, Xuhui Luo, Nuo Han, Ziyang Hu, Zitong Lin |
| Journal | dentomaxillofacial radiology |
| Year | 2026 |
| DOI |
10.1093/dmfr/twag035
|