Diagnostic Performance of a Novel Wireless Direct-Conversion Intraoral Sensor Versus CMOS and PSP Systems for Proximal Caries Detection: An Ex Vivo Study

Clicks: 3
ID: 314461
2026
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Emerging

Ranked #43 of 46 articles by views in dentomaxillofacial radiology

Most read Least read

Bar heights use a square-root scale.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
Abstract Objectives To compare the diagnostic accuracy of a new wireless direct-conversion digital intraoral sensor to current indirect conversion, complementary metal-oxide semiconductor (CMOS) and photostimulable phosphor (PSP) sensors for detecting proximal caries. Methods The sample in this ex vivo study were 28 human premolars with 21 micro-CT verified non-cavitated proximal dental carious lesions. Images were obtained using three digital intraoral systems: a wireless direct-conversion sensor, a wired indirect conversion CMOS sensor, and a PSP sensor. A custom phantom was constructed providing a limited dental arch segment comprising a molar, first premolar, and canine. The premolar was substituted and the phantom imaged using each intraoral system with (filtered) and without (raw) the application of a manufacturer-recommended enhancements, providing 168 images. Eight oral radiologists assessed the images using a five-point scale. Receiver operating characteristic analysis was performed to calculate sensitivity, specificity and the area under the curve (AUC) and compared across modalities using two-way analysis of variance (ANOVA) with post hoc Tukey’s test at an a priori level of significance of 0.05. Results The AUC ranged from 0.58 to 0.73 (sensitivity range: 0.21—0.68, specificity range: 0.69—0.92). The wireless direct-conversion sensor (0.72 ± 0.06) showed diagnostic performance comparable to CMOS (0.72 ± 0.06) and superior to PSP (0.58 ± 0.07) for raw images (p < 0.001). The direct-conversion sensor without image enhancement (0.68 ± 0.18) demonstrated the highest sensitivity, whereas the PSP without image enhancement (0.21 ± 0.09) had the lowest sensitivity. Application of image enhancement improved PSP performance (0.48 ± 0.16), approximating that of the other systems. Conclusions Based on this ex vivo study, the wireless direct-conversion and CMOS sensors demonstrated similar diagnostic performance, superior to PSP in unfiltered conditions. These findings should be interpreted in light of the ex vivo design and modest sample size.
Reference Key
openalex_W7162004961 Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors Matheus Barros‐Costa, Gustavo Machado Santaella, Gabriela Gavilán Hadid, Christiano Oliveira-Santos, Deborah Queiroz Freitas, Francisco Carlos Groppo, William C Scarfe
Journal dentomaxillofacial radiology
Year 2026
DOI
10.1093/dmfr/twag031