Magnetic resonance imaging-induced artefact mimicking torsade de pointes in a patient with an implantable cardiac monitor: a case report

Clicks: 3
ID: 313649
2026
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Star

Ranked #139 of 261 articles by views in European Heart Journal - Case Reports

Most read Least read

Bar heights use a square-root scale. Only the 120 most-read articles are drawn; the journal has 261 in total.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
Abstract Background Magnetic resonance imaging (MRI) is increasingly performed in patients with cardiac implantable electronic devices. Although generally considered safe under specific protocols, electromagnetic interference may still lead to clinically misleading recordings. Case summary A 76-year-old woman with a history of mitral valve replacement and paroxysmal atrial fibrillation underwent implantation of an implantable cardiac monitor (Abbott Confirm Rx™) for evaluation of recurrent dizziness. During brain MRI at another hospital (1.5-T GE SIGNA Creator), the device captured high-amplitude polymorphic signals resembling torsade de pointes, whereas a simultaneous 12-lead electrocardiogram confirmed sinus rhythm with a normal QT interval. A repeat MRI at our institution using a 3.0-T Siemens MAGNETOM Lumina scanner with the same time-of-flight magnetic resonance angiography sequence did not reproduce the artefact. Phantom testing on the outside hospital’s scanner also failed to demonstrate the phenomenon. However, the artefact was reproducibly observed during a follow-up MRI performed 6 months later using the same scanner. No similar findings were observed in other patients with implantable cardiac monitors scanned using the same system. Discussion MRI-related artefacts in implantable loop recorders have been previously reported. The present case demonstrates a reproducible torsade de pointes-like morphology specifically associated with the prescan phase on a particular MRI system, suggesting a device–scanner–patient interaction rather than true arrhythmia.
Reference Key
openalex_W7161041615 Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors Yasuyuki Toyama, Yusuke Taniguchi, Mayu Morisue, Y. Matsuo
Journal European Heart Journal - Case Reports
Year 2026
DOI
10.1093/ehjcr/ytag360
URL
Keywords Keywords not found

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.