Square wave voltammetry at carbon paste electrode modified with surfactant for alpha tocopheryl acetate determination in cosmetics
Article Quality & Performance Metrics
Readership in this journal
EmergingRanked #1 of 6 articles by views in potravinarstvo
Bar heights use a square-root scale.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
Abstract
The aim of this study was describe electrochemical properties of a carbon paste electrode (CPE) bulk modified with 30% (w/w) surfactant sodium dodecyl sulphate (CPE/SDS) and demonstrates its application in the determination of α‑tocopheryl acetate (α‑TAc), known as vitamin E acetate, in selected cosmetic products, especially body creams. In addition to anionic SDS, cationic hexadecylpyridinium chloride monohydrate (CPC) was also tested as possible modifier. It was found that selection of surfactant type and its content significantly affect an electrical conductivity and mechanical stability of these heterogeneous electroanalytical sensors in pure organic solvents. Under this study, it was found that CPC is a totally inappropriate mediator due to very high backgroundcurrent. Together with other lipophilic vitamins characterized by antioxidant activity (dominantly retionoids), this completely synthetic substance is widely used as significant cosmetic additive due its preservative properties. Monitoring of its content in cosmetic products is usually performed by high‑performance liquid chromatography (HPLC) with UV detection. This standard analytical protocol is always burdened with the complex and time‑consuming preparation of the sample before analysis. For that reason, robust and simple electroanalytical method based on anodic oxidation of the α‑TAc at CPE/SDS by square wave voltammetry (SWV) performed in pure organic electrolyte (99.8% acetonitrile containing 0.1 mol·L‑1 LiClO4) was developed. Moreover, simple dissolution of sample in supporting electrolyte using ultrasonic bath and subsequent filtering through a stacked filter included all the necessary procedures for sample preparation. The linear range from 0.1 to 1.2 mmol·L‑1and limit of detection 37 µmol·L‑1 were found at pulse amplitude 10 mV and frequency 10 Hz as optimum. In analysis of selected cosmetics, the developed electroanalytical method was not validated using comparison with standard HPLC. At least, the recovery was verified by analysis of model sample and value 95.8% was calculated.
| Reference Key |
abckova2018squarepotravinarstvo
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | Žabčíková, Simona;Nallbani, Arlinda;Sýs, Milan;Mikysek, Tomáš;Červenka, Libor; |
| Journal | potravinarstvo |
| Year | 2018 |
| DOI |
DOI not found
|
| URL | |
| Keywords | Keywords not found |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
ZincirX productsfrom the team behind Journament
-
PatentYazar
Patent drafts for the EPO, USPTO and TürkPatent
From inventor notes to a DOCX draft: description, claims, abstract. The machine drafts, the attorney signs.
See how it works (opens in a new tab) -
Projenator
Have a project idea? Score and plan it with AI
AI scores your idea in five dimensions and builds a phased plan. Your first project is free, no card needed.
Start your first project (opens in a new tab)
Cookies
We use strictly necessary cookies to run the site and keep you signed in. With your permission we would also use Google Analytics to see how the site is used, and Google AdSense to show ads on journal and article pages. Both stay off unless you accept, and you can change your mind at any time. How we use cookies · KVKK notice (Türkiye)
Cookie settings
Comments
No comments yet. Be the first to comment on this article.