Comparison of the novel ARCHITECT procalcitonin assay with established procalcitonin assay systems
Clicks: 126
ID: 277441
2018
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
Steady Performance
30.0
/100
126 views
39 readers
AI Quality Assessment
Not analyzed
Readership in this journal
SteadyRanked #5 of 5 articles by views in Practical laboratory medicine
Most read
Least read
Bar heights use a square-root scale.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
Aims: This study assessed the performance of a new fully automated immunoassay, ARCHITECT B.R.A.H.M.S procalcitonin (PCT), comparing the results with other commercial assays on routine clinical specimens. Methods: At nine sites from eight countries, precision analysis was carried out on controls by ANOVA. Threshold and linearity were verified according to standard procedures. Comparison of ARCHITECT B.R.A.H.M.S PCT with the Cobas®, LIAISON®, VIDAS® and Kryptor® PCT assays was evaluated using Passing-Bablok and Deming regression analyses. Results: The within-laboratory standard deviation and %CV across all sites ranged from 0.005 to 0.008 and 2.7 to 4.1; 0.040 to 0.212 and 2.1 to 11.7; 1.628 to 4.191 and 2.5–6.3 for the three control levels, respectively. The mean slope (linearity analysis) across all sites ranged from 0.85 to 1.03, with a mean y-intercept ranging from –6.15 to + 1.71 and a correlation coefficient ranging from 0.94 to 1.00. The LoB, LoD, and LoQ claims were verified. Deming regression analysis of 1116 plasma or serum samples with PCT results detected across a dynamic assay range of 0.02–100 μg/l using the ARCHITECT B.R.A.H.M.S PCT assay yielded results of r = 0.989 vs. Roche Cobas®, r = 0.986 vs Kryptor® B.R.A.H.M.S, r = 0.987 vs BioMèrieux VIDAS® and r = 0.972 vs. Diasorin LIAISON®, respectively. Concordance at cut-offs of 0.25 μg/l and 0.50 μg/l were 96.9% and 98.1% with Roche Cobas®, 95.4% and 96.1% with B.R.A.H.M.S Kryptor®, 93.8% and 98.4% with BioMèrieux VIDAS®, and 92.7% and 93.9% with Diasorin LIAISON®. Conclusions: Compared with other assays, ARCHITECT B.R.A.H.M.S PCT offers excellent precision and low-end sensitivity.
| Reference Key |
soh2018comparisonpractical
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | Soh, A.;Binder, L.;Clough, M.;Hernandez, M. Hernandez;Lefèvre, G.;Mostert, K.;Nguyen, T.B.;Otte, K.-M.;Portakal, O.;Sandri, M.S.;Yen, J.L.;Huang, J.;Beshiri, A.; |
| Journal | Practical laboratory medicine |
| Year | 2018 |
| DOI |
DOI not found
|
| URL | |
| Keywords |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.