analysis of carbon fiber reinforced peek hinge mechanism articulation components in a rotating hinge knee design: a comparison of in vitro and retrieval findings

Clicks: 203
ID: 242515
2016
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Steady

Ranked #389 of 1,324 articles by views in spectrochimica acta - part a: molecular and biomolecular spectroscopy

Most read Least read

Bar heights use a square-root scale. Only the 120 most-read articles are drawn; the journal has 1,324 in total.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
Carbon fiber reinforced poly-ether-ether-ketone (CFR-PEEK) represents a promising alternative material for bushings in total knee replacements, after early clinical failures of polyethylene in this application. The objective of the present study was to evaluate the damage modes and the extent of damage observed on CFR-PEEK hinge mechanism articulation components after in vivo service in a rotating hinge knee (RHK) system and to compare the results with corresponding components subjected to in vitro wear tests. Key question was if there were any similarities or differences between in vivo and in vitro damage characteristics. Twelve retrieved RHK systems after an average of 34.9 months in vivo underwent wear damage analysis with focus on the four integrated CFR-PEEK components and distinction between different damage modes and classification with a scoring system. The analysis included visual examination, scanning electron microscopy, and energy dispersive X-ray spectroscopy, as well as surface roughness and profile measurements. The main wear damage modes were comparable between retrieved and in vitro specimens (n=3), whereby the size of affected area on the retrieved components showed a higher variation. Overall, the retrieved specimens seemed to be slightly heavier damaged which was probably attributable to the more complex loading and kinematic conditions in vivo.
Reference Key
schierjott2016biomedanalysis Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors ;Ronja A. Schierjott;Alexander Giurea;Hans-Joachim Neuhaus;Jens Schwiesau;Andreas M. Pfaff;Sandra Utzschneider;Gianluca Tozzi;Thomas M. Grupp
Journal spectrochimica acta - part a: molecular and biomolecular spectroscopy
Year 2016
DOI
10.1155/2016/7032830
URL
Keywords

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.