a new model to study fatigue in dental implants based on probabilistic finite elements and cumulative damage model

Clicks: 255
ID: 236412
2017
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Emerging

Ranked #7 of 31 articles by views in water-rock interaction - proceedings of the 13th international conference on water-rock interaction, wri-13

Most read Least read

Bar heights use a square-root scale.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
The aim of this study was to predict the fatigue life of two different connections of a dental implant as in load transfer to bone. Two three-dimensional models were created and assembled. All models were subjected to a natural masticatory force of 118 N in the angle of 75° to the occlusal plane. All degrees of freedom in the inferior border of the cortical bone were restrained, and the mesial and distal borders of the end of the bone section were constrained. Fatigue material data and loads were assumed as random variables. Maximum principal stresses on bone were evaluated. Then, the probability of failure was obtained by the probabilistic approach. The maximum principal stress distribution predicted in the cortical and trabecular bone is 32 MPa for external connection and 39 MPa for internal connection. A mean life of 103 and 210 million cycles were obtained for external and internal connection, respectively. Probability cumulative function was also evaluated for both connection types. This stochastic model employs a cumulative damage model and probabilistic finite element method. This methodology allows the possibility of measured uncertainties and has a good precision on the results.
Reference Key
prados-privado2017applieda Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors ;María Prados-Privado;José Antonio Bea;Rosa Rojo;Sérgio A. Gehrke;José Luis Calvo-Guirado;Juan Carlos Prados-Frutos
Journal water-rock interaction - proceedings of the 13th international conference on water-rock interaction, wri-13
Year 2017
DOI
10.1155/2017/3726361
URL
Keywords

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.