assessment of an automatic robotic arm for dispensing of chemotherapy in a 2500-bed medical center
Clicks: 252
ID: 235037
2013
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
Steady Performance
30.0
/100
252 views
52 readers
AI Quality Assessment
Not analyzed
Readership in this journal
SteadyRanked #84 of 246 articles by views in Bioorganic & medicinal chemistry letters
Most read
Least read
Bar heights use a square-root scale. Only the 120 most-read articles are drawn; the journal has 246 in total.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
Automation has long been awaited in parenteral drug dispensing. Pharmacists can benefit much in theory from a good automated device to handle the hazardous drugs used in chemotherapy. This paper describes the performance of the first chemotherapy-dispensing robot in the oncology pharmacy of a 2500-bed medical center. The objective of this paper is two-fold: (1) to assess the robot’s performance in terms of its success rate and to summarize the causes of failure, and (2) to find out if the robot can decrease the full-time equivalents (FTEs) of the oncology pharmacy.
Methods: We used the computer-generated log from the first week of May 2010 to that of July 2010, supplemented with the pharmacists’ notes on the causes of failure, to determine the success rate and to analyze the incidences of failure. We also assessed the FTEs before and after implementing the robot.
Results: Data showed that the success rate rose slowly from 76.8% to 95.3% over the 2-month recording period. The major mechanical problems encountered were air, clamping, and waste bin problems. Manual errors, such as loading wrong drugs or syringes, also caused failures. In terms of manpower saving, CytoCare failed to decrease the number of FTE pharmacists/technicians in our oncology pharmacy practice.
Conclusion: We conclude that even though CytoCare could ease the risk of chemotherapy exposure and increase the precision of dosing, it was not able to improve the FTE pharmacists/technicians in our hospital.
| Reference Key |
chen2013journalassessment
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | ;Wen-Hwei Chen;Li-Jiuan Shen;Ru-Jiun Guan;Fe-Lin Lin Wu |
| Journal | Bioorganic & medicinal chemistry letters |
| Year | 2013 |
| DOI |
10.1016/j.jfma.2011.11.026
|
| URL | |
| Keywords |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments