a validated rp-hplc method for the determination of recombinant human insulin in bulk and pharmaceutical dosage form

Clicks: 134
ID: 228198
2010
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Emerging

Ranked #154 of 235 articles by views in Journal of pharmaceutical and biomedical analysis

Most read Least read

Bar heights use a square-root scale. Only the 120 most-read articles are drawn; the journal has 235 in total.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
A modified RP-HPLC method was developed for the quantitative determination of recombinant human insulin in bulk and pharmaceutical dosage form with reduced retention time. Study of the effects of the column temperature, pH of the mobile phase and presence of vial additives (phenol and m-cresol), or impurities (A-21 Disamido) on the accuracy of the assay were assessed. Separation was achieved using a Hypersil BDS C-18 column and the mobile phase was composed of solution A (aqueous solution of 28.3 anhydrous Na2SO4g/L, pH 2.3) and solution B (28.5 g anhydrous Na2SO4 g/L in 50:50 mixture of water and acetonitrile, pH 2.3) in a ratio 48:52 (v/v) at 45–50 °C. The column temperature was 40 °C, the flow rate was 1 mL/min and detection was performed at 216 nm. The procedures were validated according to international conference on harmonization (ICH) guidelines. Recovery study was done applying standard addition technique for further validation of the procedure. The retention time of recombinant human insulin was 19.7 min as compared to 29 min obtained by the reference method. Analytical conditions fluctuations or presence of vial additives or impurities did not show any significant effect on the accuracy of the method. The prepared standard insulin solution in 0.01 N HCl was found to be stable for 5 days. Statistical comparison showed no significant difference between the described method and reference method regarding the accuracy and precision. The modified method can be applied for routine quality control applications for determination of recombinant human insulin.
Reference Key
moussa2010e-journala Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors ;B. A. Moussa;F. Farouk;H. M. E. Azzazy
Journal Journal of pharmaceutical and biomedical analysis
Year 2010
DOI
10.1155/2010/147824
URL
Keywords

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.