An Embedded Device for Real-Time Noninvasive Intracranial Pressure Estimation.
Clicks: 287
ID: 22349
2018
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
Steady Performance
65.1
/100
287 views
198 readers
Trending
AI Quality Assessment
Not analyzed
Readership in this journal
SteadyRanked #1 of 3 articles by views in acta neurochirurgica supplement
Most read
Least read
Bar heights use a square-root scale.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
The monitoring of intracranial pressure (ICP) is indicated for diagnosing and guiding therapy in many neurological conditions. Current monitoring methods, however, are highly invasive, limiting their use to the most critically ill patients only. Our goal is to develop and test an embedded device that performs all necessary mathematical operations in real-time for noninvasive ICP (nICP) estimation based on a previously developed model-based approach that uses cerebral blood flow velocity (CBFV) and arterial blood pressure (ABP) waveforms.The nICP estimation algorithm along with the required preprocessing steps were implemented on an NXP LPC4337 microcontroller unit (MCU). A prototype device using the MCU was also developed, complete with display, recording functionality, and peripheral interfaces for ABP and CBFV monitoring hardware.The device produces an estimate of mean ICP once per minute and performs the necessary computations in 410 ms, on average. Real-time nICP estimates differed from the original batch-mode MATLAB implementation of theestimation algorithm by 0.63 mmHg (root-mean-square error).We have demonstrated that real-time nICP estimation is possible on a microprocessor platform, which offers the advantages of low cost, small size, and product modularity over a general-purpose computer. These attributes take a step toward the goal of real-time nICP estimation at the patient's bedside in a variety of clinical settings.
| Reference Key |
matthews2018anacta
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | Matthews, Jonathan M;Fanelli, Andrea;Heldt, Thomas; |
| Journal | acta neurochirurgica supplement |
| Year | 2018 |
| DOI |
10.1007/978-3-319-65798-1_18
|