Abstract
In the current study, an extended 3-pore model (TPM) is presented and applied to the problem of optimizing automated peritoneal dialysis (APD) with regard to osmotic water transport (UF), small/middle-molecule clearance, and glucose absorption. Methods: Simulations were performed for either intermittent APD (IPD) or tidal APD (TPD). IPD was simulated for fill and drain volumes of 2 L, whereas TPD was simulated using a tidal volume of 0.5 L, 1 L, or 1.5 L with full drains and subsequent fills (2 L) occurring after every fifth dwell. A total of 25 cycles for a large number of different dialysate flow rates (DFR) were simulated using 3 different glucose concentrations (1.36%, 2.27%, and 3.86%) and 3 different peritoneal transport types: slow (peritoneal equilibrium test D/Pcrea < 0.6), fast (peritoneal equilibrium test D/Pcrea > 0.8), and average. Solute clearance and UF were simulated to occur during the entire dwell, including both fill and drain periods. Results: It is demonstrated that DFRs exceeding ∼ 3 L/h are of little benefit both for UF and small-solute transport, whereas middle-molecule clearance is enhanced at higher DFRs. The simulations predict that large reductions (> 20%) in glucose absorption are possible by using moderately higher DFRs than a standard 6 × 2 L prescription and by using shorter optimized “bi-modal” APD regimens that alternate between a glucose-free solution and a glucose-containing solution. Discussion: Reductions in glucose absorption appear to be significant with the proposed regimens for APD; however, further research is needed to assess the feasibility and safety of these regimens.
Reference Key
berg2017kidneyoptimizing Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors ;Carl M. Öberg;Bengt Rippe
Journal Soft matter
Year 2017
DOI