comparison of central corneal thickness measured by anterior segment optical coherence tomography, pentacam and ultrasonic pachymetry

Clicks: 272
ID: 212948
2013
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This article has not been analysed, so there is no overall score — reader engagement is measured and shown alongside.
AI Quality Assessment
Not analyzed
Readership in this journal
Steady

Ranked #52 of 133 articles by views in journal of fungi

Most read Least read

Bar heights use a square-root scale. Only the 120 most-read articles are drawn; the journal has 133 in total.

Mint this article as an NFT
Not yet minted

Create a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.

5 SUSD one-off · no wallet required
Abstract
AIM: To investigate and analyze the differences in central corneal thickness(CCT)measurement by anterior segment optical coherence tomography(AS-OCT), pentacam rotating Scheimpflug camera and ultrasonic pachymetry(USP).

METHODS: The retrospective study measured CCT of 43 patients(86 eyes)performed excimer laser myopia corrective surgery by the same examiner using RTVue AS-OCT with an anterior segment adaptor, pentacam rotating Scheimpflug camera, and PACSCAN300P USP. Data of three groups were compared and statistically analyzed by pair t test.

RESULTS: The mean CCT(±SD)measured by AS-OCT,pentacam and USP were 535±52, 550±52, 550±51μm, respectively. Paired-sample t test showed that the results of AS-OCT, pentacam and USP were statistically different. There were no statistically significant differences in the mean CCT between the pentacam and USP group. Significant correlations of measurement result were found between pentacam and USP(r=0.998, P<0.01), FD-OCT and USP(r=0.993, P<0.01), and AS-OCT and pentacam(r=0.991, P<0.01).

CONCLUSION:RTVue AS-OCT significantly under estimates the CCT, while the measurement results by pentacam are similar to those by USP which is regarded as the golden standard.

Reference Key
zhao2013guojicomparison Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors ;Wei Zhao;Ting Wu;Ze-Hong Dong;Jie Feng;Yu-Feng Ren;Yu-Sheng Wang
Journal journal of fungi
Year 2013
DOI
10.3980/j.issn.1672-5123.2013.12.59
URL
Keywords Keywords not found

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org

No comments yet. Be the first to comment on this article.