ecg-based detection of early myocardial ischemia in a computational model: impact of additional electrodes, optimal placement, and a new feature for st deviation
Clicks: 260
ID: 199570
2015
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
Steady Performance
65.4
/100
260 views
213 readers
Trending
AI Quality Assessment
Not analyzed
Readership in this journal
SteadyRanked #92 of 1,324 articles by views in spectrochimica acta - part a: molecular and biomolecular spectroscopy
Most read
Least read
Bar heights use a square-root scale. Only the 120 most-read articles are drawn; the journal has 1,324 in total.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
In case of chest pain, immediate diagnosis of myocardial ischemia is required to respond with an appropriate treatment. The diagnostic capability of the electrocardiogram (ECG), however, is strongly limited for ischemic events that do not lead to ST elevation. This computational study investigates the potential of different electrode setups in detecting early ischemia at 10 minutes after onset: standard 3-channel and 12-lead ECG as well as body surface potential maps (BSPMs). Further, it was assessed if an additional ECG electrode with optimized position or the right-sided Wilson leads can improve sensitivity of the standard 12-lead ECG. To this end, a simulation study was performed for 765 different locations and sizes of ischemia in the left ventricle. Improvements by adding a single, subject specifically optimized electrode were similar to those of the BSPM: 2–11% increased detection rate depending on the desired specificity. Adding right-sided Wilson leads had negligible effect. Absence of ST deviation could not be related to specific locations of the ischemic region or its transmurality. As alternative to the ST time integral as a feature of ST deviation, the K point deviation was introduced: the baseline deviation at the minimum of the ST-segment envelope signal, which increased 12-lead detection rate by 7% for a reasonable threshold.
| Reference Key |
loewe2015biomedecg-based
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | ;Axel Loewe;Walther H. W. Schulze;Yuan Jiang;Mathias Wilhelms;Armin Luik;Olaf Dössel;Gunnar Seemann |
| Journal | spectrochimica acta - part a: molecular and biomolecular spectroscopy |
| Year | 2015 |
| DOI |
10.1155/2015/530352
|
| URL | |
| Keywords |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.