Instant velocity and consistency of emitted cloud change by the different levels of canister filling with Metered Dose Inhalers (MDIs), but not with Soft Mist Inhalers (SMIs): a bench study
Clicks: 333
ID: 19405
2017
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
Steady Performance
64.1
/100
333 views
236 readers
Trending
AI Quality Assessment
Not analyzed
Readership in this journal
SteadyRanked #2 of 3 articles by views in multidisciplinary respiratory medicine
Most read
Least read
Bar heights use a square-root scale.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
Abstract Background Inhalation is the preferred route for respiratory drug delivery, but several factors contribute to the variability of the respirable dose fraction. Instant velocity and the dynamic characteristics of the droplet cloud represent crucial factors. Aim was to measure and compare the instant velocity and the consistency of emitted cloud from five different MDIs (A - Salbutamol sulphate 100mcg, GSK; B - Salbutamol sulphate 100mcg, Valeas; C - Salmeterol xinafoate/Fluticasone propionate 25/125mcg, GSK; D - Formoterol fumarate/Bechlomethasone propionate 6/100mcg, Chiesi; E - Formoterol fumarate/Fluticasone dipropionate 5/125mcg, Mundipharma) and one SMI (Tiotropium bromide 5mcg, Boehringer Ingelheim), at different distance from the nozzle and canister filling. Methods Measurements were made at 90, 50, and 10% of canister filling, and at 5, 10, and 20 cm from the nozzle, for a total of 972 puffs. A high speed video photography protocol was adopted and high speed cameras (1.200 frames/sec.) were used. Data were acquired by means of specialized softwares. Temperature, humidity, and vibrations occurrence were strictly controlled during measurements. Statistics: Anova and p < 0.05 were accepted as the minimum significance level. Results MDIs generated different Instant velocities: MDI B generated the highest, while MDI A the lowest. As expected, velocity decreased in proportion to the distance from the nozzle. Except with MDI C, instant velocity decreased significantly over the first 50% of canister emptying, but dropped by >33% at 90% of emptying with all other MDIs (p < 0–037; p < 0.001; p < 0.005, and p < 0.001, respectively). Instant velocity was extremely lower (p < 0.001) and constant for all levels of canister filling (p = ns) with SMI. All MDIs had a very fast jet phase, ranging 0.01–0.03 s at 10 cm, and 0.03–0.05 s at 20 cm from the nozzle, without any significant difference from each other (p = ns). MDIs generated a cloud similarly tight (p = ns) at 10 and 20 cm from the nozzle, while it was extremely wider and constant with the SMI (p = 0–001). Also the cloud turbulence was minimized during the SMI emission. Discussion and Conclusions MDIs are characterized by a substantial variability in both their instant velocity and consistency of the emitted cloud at different levels of canister filling. SMI generates a much slower soft mist cloud which is constantly homogeneous and independent of canister emptying. These peculiarities assessed at bench are suggesting a higher dose consistency and a much more effective therapeutic performance also in real life.
| Reference Key |
negro2017instantmultidisciplinary
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | Negro, Roberto W. Dal;Longo, Pietro;Ziani, Orestepaolo Villanis;Bonadiman, Luca;Turco, Paola; |
| Journal | multidisciplinary respiratory medicine |
| Year | 2017 |
| DOI |
DOI not found
|
| URL | |
| Keywords | Keywords not found |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.