development and validation of immunoenzyme test-system for determination of 25-hydroxyvitamin d in blood serum

Clicks: 154
ID: 190748
2016
Article Quality & Performance Metrics
Overall Quality Improving Quality
0.0 /100
Combines engagement data with AI-assessed academic quality
AI Quality Assessment
Not analyzed
Abstract
The present study was aimed at designing of immunoenzyme test-system for competitive determination of 25OHD in blood serum using polyclonal antibodies against 25OHD and biotin-streptavidin visualization technique. The validation parameters of test-system were determined. For development of test-system we used the preparation of specific antibodies against 25OHD with the titer 1:3000. The optimal concentration of competing agent (25OHD3-LC-biotin) was found to be equal to 5 ng/ml. The standard calibration curve was built by using a set of seven calibrators. For the validation of test-system the following parameters were established: limit of detection, quantitative limit, cross-reactivity, intra- and inter-system coefficients of variation (Intra.CV, Inter.CV) and “matrix effect”. Constructed immunoenzyme test-system can be used successfully for 25OHD determination in blood serum samples.
Reference Key
2016biotechnologiadevelopment Use this key to autocite in the manuscript while using SciMatic Manuscript Manager or Thesis Manager
Authors ;A. O. ;I. O. ; M. M.
Journal third world quarterly
Year 2016
DOI
10.15407/biotech9.02.028
URL
Keywords

Citations

No citations found. To add a citation, contact the admin at info@scimatic.org