a new method of estimating intracranial elastance
Clicks: 88
ID: 178076
2014
Article Quality & Performance Metrics
Overall Quality
Not rated
Combines reader engagement with the AI quality analysis. This
article has not been analysed, so there is no overall score —
reader engagement is measured and shown alongside.
Reader Engagement
Steady Performance
25.5
/100
88 views
17 readers
AI Quality Assessment
Not analyzed
Readership in this journal
SteadyRanked #44 of 45 articles by views in advances in materials science and engineering
Most read
Least read
Bar heights use a square-root scale.
Mint this article as an NFT
Not yet mintedCreate a permanent, verifiable on-chain record of this article on the Scimatic Network. The NFT is held in your Journament account, and you can withdraw it to your own wallet at any time.
5
SUSD
one-off · no wallet required
Abstract
Objective: Current methods of calculating Intracranial Elastance Index (IEI) depend from CSF pulse-wave, whose shape may deeply change during ICP rising.
The main aim of this study was to evaluate the reliability and specificity of a novel method to calculate IEI (method C), based on the integral of the CSF pulse-wave area.
Method: Twenty ventricular infusion-tests of patients with idiopathic NPH were re-evaluated. We have compared method C with the most widely used methods to calculate IEI: a modified Szewczykowski method (diastolic ICP against CSF pulse-wave amplitude-method A) and a modified Czosnyka method (diastolic ICP against the fundamental harmonic-method B). R-squared (R2) was calculated for each test. Means were compared through ANOVA and t-test.
Results: Mean R2 values for methods A, B and C were 0.91 ± 0.06, 0.9 ± 0.06 and 0.96 ± 0.03, respectively. Mean R2 values obtained through method A vs C and through method B vs C were significantly different (p = .006 and p = .001, respectively), while values obtained through method A vs B were not (p = 1). Analysis of ICP tracks demonstrated that 9 patients showed no different shape of the ICP wave during the infusion test, while the remaining 11 did. The mean R2 values obtained through method A vs C and through method B vs C were significantly different (p < .001 for both) for patients showing a different shape of the ICP wave during the infusion test.
Conclusions: Method C seems to be the most reliable method to calculate IEI, as it is independent from CSF pulse wave modifications.
| Reference Key |
md2014interdisciplinarya
Use this key to autocite in the manuscript while using
SciMatic Manuscript Manager or Thesis Manager
|
|---|---|
| Authors | ;Carmelo Anile, MD;Pasquale De Bonis, PhD;Annunziato Mangiola, MD;Stefano Mannino, PhD;Pietro Santini |
| Journal | advances in materials science and engineering |
| Year | 2014 |
| DOI |
10.1016/j.inat.2014.03.002
|
| URL | |
| Keywords |
Citations
No citations found. To add a citation, contact the admin at info@scimatic.org
Comments
No comments yet. Be the first to comment on this article.